21
4.10.1
Configuration of the test parameters for RAMP#1 mode
a.
By cyclically pressing the
SET/DISPLAY MEM
key, it is possible to select the following
parameters:
Minimum value of the insulation (MIN) in a range between 0.01M
- MAX
Maximum value of the insulation (MAX) in a range between (MIN) - 10T
Value of time interval ΔT, in which the test voltage passes from 0V to the set value V1
Value of the test voltage V1 in a range between 100V - 5000V (can be set in steps of 25V
up to 1000V and in steps of 50V from 1000V to 5000V)
Value of time T1 in which the test voltage V1 is applied, in a range between
(00min:05sec) - (99min:59 sec)
Number N of cycles in which the ramp is repeated, in a range between 1 - 255 (N=1
indicates 1 only test without repetitions)
b.
Press the arrow keys to change the value of the parameters (a long pressing of the keys
allows a rapid change of the values)
c.
Press the
MODE/CLEAR SMOOTH
key to remove the limit values set for the parameters. The
display shows the message "no" to indicate that there is no limit set.
When setting the value of the time interval ΔT, of the test voltage V1, of time T1 and of the
number N of cycles, by pressing the
MODE/CLEAR SMOOTH
key, the default values are set
(respectively 30sec, 1000V, 1min, 1 cycle)
d.
After setting, press the
SAVE
key to save the changes made or press the
ESC/RECALL MEM
key to exit without saving and return to the previous screen
4.10.2
Configuration of the test parameters for RAMP#2 mode
a.
By cyclically pressing the
SET/DISPLAY MEM
key, it is possible to select the following
parameters:
Minimum value of the insulation (MIN) in a range between 0.01M
- MAX
maximum value of the insulation (MAX) in a range between (MIN) - 10T
Value of the starting voltage V0 in a range between 100V - 1000V (can be set in steps of
25V)
Value of time T0 in which the test voltage V0 is applied, in a range between
(00min:05sec) - (99min:59 sec)
Value of time interval ΔT in which the test voltage passes from the set value V0 to the
set value V1, in a range between (00min:05sec) - (99min:59 sec);
Value of the test voltage V1 in a range between 1000V - 5000V (can be set in steps of
50V)
Number N of cycles in which the ramp is repeated, in a range between 1 - 255 (N=1
indicates 1 only test without repetitions).
b.
Press the arrow keys to change the value of the parameters (a long pressing of the keys
allows a rapid change of the values).
c.
Press the
MODE/CLEAR SMOOTH
key to remove the limit values set for the parameters. The
display shows the message "no" to indicate that there is no limit set.
When setting the value of voltage V0, of time T0, of time interval ΔT, of voltage V1 and of
the number N of cycles, by pressing the
MODE/CLEAR SMOOTH
key, the default values are
set (respectively 100V, 15sec, 30sec, 1000V, 1 cycle).
d.
After setting, press the
SAVE
key to save the changes made or press the
ESC/RECALL MEM
key to exit without saving and return to the previous screen.