3.2.1.12 Energy Digit Reset Count (Rollover Count)
This screen enables the user for setting maximum energy count after which energy will rollover
to zero. This rollover count values are 7, 8 and 9 Digits.
Touching radio button in front of particular option will select that
option.
Touch on “ key” will conrm the setting.
Touching the “ key” will keep the old selected setting and will
return to previous menu.
OK
OK
BACK
Note :-
1) If Energy Resolution is set to MW & energy digit reset count is set to 9, Energy
screen on display will show “-------” i.e energy overow when energy crosses the
8 digit count.
LINE-NEUTRAL VOLTAGE
VL2
VL2
VL2
LINE-NEUTRAL VOLTAGE
VL2
VL2
VL2
LINE-NEUTRAL VOLTAGE
VL2
VL2
VL2
CT SECONDARY
1 AMPERE
LINE-NEUTRAL VOLTAGE
VL2
VL2
VL2
LINE-NEUTRAL VOLTAGE
VL2
VL2
VL2
LINE-NEUTRAL VOLTAGE
VL2
VL2
VL2
ENERGY DIGIT RESET COUNT
7 DIGITS
9 DIGITS
8 DIGITS
OK
BACK
3.2.1.14 Meter Version
3.2.1.13 Energy Update Rate
This screen allows user to enter energy update rate in minutes. After entering particular value in
minutes, the energy will be updated on modbus location from 30145 to 30165 of 3X register and
40145 to 40165 of 4X register as per value that user has entered.
INVALID VALUE
ENERGY UPDATE RATE
BACK
OK
OK
The valid range of number of poles is from
1 to 60
minutes.
If a value outside the range is entered, It will display “INVALID
VALUE” followed by the correct range of parameter.
Touch on “ key” will conrm the setting and will return to
previous menu. Touching the “ key” will keep the old selected
setting.
If Energy Rate is set to 2 then energy will get stored after 2 minutes
on the modbus.
The meter version is available on the SYSTEM PARAMETERS screen as read only.