![Rion NA-42 Скачать руководство пользователя страница 82](http://html1.mh-extra.com/html/rion/na-42/na-42_instruction-manual_1463440082.webp)
72
Serial Interface
MAX n
Set
L
max
(sound level maximum) function to On/Off
n = 0
Maximum hold OFF
n = 1
Maximum hold ON
MAX? Get
L
max
function status
Output data format
p<
EOT
><
CR
><
LF
>
p = 0
L
max
OFF
p = 1
L
max
active
PEK n
Set
L
peak
(peak sound pressure level) function to On/
Off
n = 0
L
peak
OFF
n = 1
L
peak
ON
PEK?
Get
L
peak
function status
Output data format
p<
EOT>
<
CR
><
LF
>
p = 0
L
peak
OFF
p = 1
L
peak
active
RST
Reset hold value
This command is active during
L
max
,
L
peak
, and compara-
tor operation.
CMP n
Set comparator function to On/Off
n = 0
Comparator OFF
n = 1
Comparator ON
CMP ?
Get comparator status
Output data format
p<
EOT
><
CR
><
LF
>
p = 0
Comparator OFF
p = 1
Comparator active