Serial Interface
Error Processing
In order to ensure correct data exchange between the NA-42 and the computer,
the rules described above must be observed. If an error occurs, the following
steps should be taken.
The computer has sent RMT1
**
<
CR
><
LF
> but the REMOTE indica-
tor on the NA-42 does not light.
Send RMT1
**
<
CR
><
LF
> again after about 4 seconds. Repeat this
2 or 3 times. If the REMOTE indicator still does not light, one of
the following conditions may exist:
** stands for ID number.
·
Transfer parameters do not match. (Check settings made with menu
screens.)
·
Interface cable is defective or not properly connected.
·
NA-42 is not powered.
·
ID number setting does not match.
When the computer has sent a wrong command or the parameter is out of
range, the NA-42 disregards that command, and also when the ID number
is other than FF, the NA-42 returns <
NAK
>READY<
CR
><
LF
> . If the NA-
42 is in receiving mode but there is no command from the computer for 4
seconds or more, the receive mode times out.
<
NAK
>:
Control code 15
H
Control Operation in Remote Mode
While the NA-42 is in remote mode, only the MENU (LIGHT) key and RESET
key are active. The MENU (LIGHT) key allows turning the display backlight
on and off, and the RESET key in remote mode serves for switching back to
local mode. This allows manual override of remote mode at any time.