59
Serial Interface
Local Mode/Remote Mode
Local
mode
In this mode, the controls on the panel of the NA-42 are used to op-
erate the unit. Immediately after being turned on, the unit is always
in local mode.
Remote
mode
In this mode, the NA-42 operates in response to commands sent
from a computer. Only the MENU (LIGHT) key and RESET key on
the NA-42 are active, and cannot be operated by all other keys. The
indication REMOTE appears on the display of the NA-42 when the
unit is in remote mode.
Switching between local mode and remote mode
The RMT command is used to switch between local mode and remote
mode.
Preparation
1. Turn power to the NA-42 and the computer off.
2. Use a generic cable to connect the serial interface connector on the
rear panel of the NA-42 to the computer.
Cable:
Generic cross-wired serial cable (null
modem)
Connector type on NA-42: 9-pin D-sub, male
1:1 connection
SERIAL
COMP OUT
±12V
±45V
PREAMP
A
COM
LOCK
UNLOCK
BATT BOX
AC OUT
DC OUT
EXT 6V
TO
FILTER
FROM
FILTER
EXT
FILTER
NA-42
Serial interface
connector
Cross-wired serial cable
(null modem)
9-pin D-sub plug,
female
Computer