![Rion NA-42 Скачать руководство пользователя страница 114](http://html1.mh-extra.com/html/rion/na-42/na-42_instruction-manual_1463440114.webp)
104
Specifi cations
Measurement functions
Sound pressure level
L
p
Sound level maximum
L
max
Maximum sound level value is held on measurement value readout.
Peak sound pressure level
L
peak
Peak sound pressure level value is held on measurement value read-
out.
Sampling frequency 250 kHz
Frequency range
1 Hz to 100 kHz (approx -2 dB at 1 Hz, approx -3 dB at 100 kHz)
Frequency weighting
A, C, FLAT
Linear operation range and applicable standards
Depending on microphone combination
(Linear operation range applies to A weighting without overload mar-
gin)
The Linearity range values given in the above table are guaranteed for
all microphones of the indicated type.
When wider linear operation range is required, microphone selection can
extend upper or lower limit by about 1 to 5 dB
*
.
* Depends on microphone type.
UC-52
UC-34P
UC-29
UC-27
UC-31
UC-53A
28 to 146 dB
10 to 112 dB
50 to 164 dB
20 to 145 dB
34 to 155 dB
28 to 145 dB
20 Hz to 8 kHz
10 Hz to 12.5 kHz
20 Hz to 100 kHz
5 Hz to 12.5 kHz
10 Hz to 35 kHz
10 Hz to 20 kHz
IEC 60651
TYPE 2
TYPE 1
TYPE 1
TYPE 1
TYPE 1
TYPE 1
C1502
C1505
C1505
C1505
C1505
JIS
C1505
Designation Linear operation range
Frequency range