9.PRODUCT SPECIFICATIONS
9−1.Standard specifications
Model
: GD-F4A-1
Detection principle
: Galvanic cell method
Target gas
: Oxygen in air
Sampling method
: Sample-drawing(flow-through type)
Signal transmission
: Sensor output direct transmission
Transmission cable
: CVVS 1.25mm
2
, 2C cable or equivalent
Transmission distance
: Max 600m with CVVS 1.25mm
2
, 2C cable
(between indicator/alarm unit and detector head)
Applicable zener barrier
: MTL 761Pac
Cable entry
: Cable gland (Applicable cable O.D. : φ8〜φ11mm)
Piping connection
: Rc 1/8, φ8-1t(Provided with coupling)
Operating temperature
: -10〜40℃
Operating humidity
: Below 90%RH (Non-condensing)
Structure
: Wall mounting/2B(50A) pole mounting
Kind of explosion proof
: Intrinsically safe(when zener barrier is used)
Explosion proof class
: ExiaⅡCT4X(when zener barrier is used)
Outer dimensions
: Approx 175(H)×140(W)×86(D)mm
Weight
: Approx 1.5kg
Painting color
: Munsel No.2.5Y9/2 half tone
* Specifications subject to change without notice.
9−2.Standard accessories
・Operating manual
・Sensor
9−3.Optional accessories
・U-bolt (for nominal diameter 2B)
・Zener barrier (MTL761Pac)