9 Product Specifications
9-1. Standard specifications
- 30 -
9
Product Specifications
9-1. Standard specifications
<
ATEX/IECEx Specifications
>
Model GD-A80-70
Detection
principle
Catalytic combustion
Mounted sensor
HW-6212
HW-6239
Gas to be
detected
Combustible gas(methane excluded),
Solvent vapor
Methane, Isobutane, H
ydrogen
Detection method
Natural diffusion type
Transmission
cable
EM CCE/F (1.25 or 2.0 mm
2
), 4-core equivalent
Cable connecting
port
Pressure proof packing gland G3/4
[Rubber seal ID] ø12 for mounted, ø13 for attached
Operating
temperatures
-10 - +100
℃
(At a constant condition)*
-40 - +70
℃
(At a constant condition)
Operating
humidities
Below 95% RH (Non-condensing)
Structure
Wall mounting type, 2B (50A) pipe attachable
Explosion-proof
structure
Flame-proof enclosures
Explosion-proof
class
Ⅱ
2G Ex db
Ⅱ
C T4 Gb (ATEX)
Ex db
Ⅱ
C T4 Gb (IECEx)
External
dimensions
Approx. 78 (W) x 163 (H) x 105 (D) mm (Fixing brackets excluded)
Weight Approx.
1.0kg
Paint color
Munsell 7. 5BG5/2
Name plate : red
*The ambient temperature range which maintain a capability of explosion proof of the label on the detector
is -40
~
+70
℃
. However, the available operating temperature range is -10
~
+100
℃
(when it exceed
+70
℃
, is exempt from explosion proof).