44
9-2. Troubles in reading
Symptoms
Causes
Actions
Incorrect reading
Reference calibration has not
been performed properly.
Perform reference calibration again in fresh air.
Presence of interference gas
If an interference gas (other than SF6 and air)
is included, correct measurement cannot be
performed by the purity meter. Remove the
interference gas and perform measurement
again.