NOTE
In a low-temperature environment, the operating time is shortened due to the battery performance
property.
At a low temperature, the response of the LCD display may get slow down.
4-8. Performing measurement
Switch the three-way switching valve to the SAMPLING side to start supplying a gas to the detector head.
Wait until replacement of gas inside the detector head is completed. The following table shows gas flow
rates and the time required for replacement at that time for reference.
When replacement of gas is completed, switch the three-way switching valve to the STOP side read the
minimum value in 5 sec after gas supplying stops.
Supply pressure
0.1 MPa
0.2 MPa
0.3 MPa
0.4 MPa
0.5 MPa
0.6 MPa
0.7 MPa
Gas flow rate
60 ml/min
90 ml/min
120 ml/min
150 ml/min
180 ml/min
210 ml/min 250 ml/min
Time required for
replacement
60
seconds
45
seconds
30
seconds
25
seconds
20
seconds
15
seconds
15
seconds
NOTE
Replacement may take extra time if air remains in the sampling hose or upstream tube.
Note that above gas flow rates and the time required for replacement are given for reference.
Supply gas
Read
indication