![Riken Keiki EC-6002 Скачать руководство пользователя страница 74](http://html1.mh-extra.com/html/riken-keiki/ec-6002/ec-6002_operating-manual_1461568074.webp)
10 Product Specifications
10-1. List of specifications
- 73 -
RM-6000
10
Product Specifications
10-1. List of specifications
<Common Specifications>
Concentration
display
Character LCD (digital and bar meter display <three colors: green, orange, and
red>)
Power display
POWER lamp on (green)
Gas alarm display
First: ALM1 lamp blinks or lights up (red)/buzzer
Second: ALM2 lamp blinks or lights up (red)/buzzer
Gas alarm pattern
Self-latching, auto-reset, or lock-in
Gas alarm contact
No-voltage contact 1a or 1b (2 step independent)
De-energized (energized at an alarm) or energized (de-energized at an alarm)
Fault alarm display
FAULT lamp blinking (orange)/Detail display/Buzzer sounds
Fault alarm pattern
Auto-reset
Fault alarm contact
No-voltage contact 1a or 1b
De-energized (energized at an alarm) or energized (de-energized at an alarm)
Contact capacity
125 VAC - 1A/30 VDC - 1A (Resistance load)
Transmission
system
Analog/digital transmission [option]
Transmission
specifications
Analog transmission: 4 - 20 mA DC (no-insulation/load resistance under 300
Ω
)/digital transmission: RS-485
Power supply
AC specification: 100 - 240 VAC ±10%, 50/60 Hz, or
DC specification: 24 VDC ±10% (21.6 - 26.4 VDC) [option]
Utility
AC output for pump power supply(AC100
~
240V
・
MAX.15VA )
DC output for pump power supply(DC24V
・
MAX.9W) [option]
Initial clear*
Approx. 25 seconds
Operating
temperatures
-10 - 50°C (at a constant condition)
Operating
humidities
Below 95% RH (Non-condensing)
Structure
Wall mounting type or rack mounting type
External
dimensions
Wall mounting type: Approx. 110 (W) x 190 (H) x 54 (D) mm (projection
portions excluded)
Rack mounting type: Approx. 110 (W) x 190 (H) x 54 (D) mm (projection
portions excluded)
Weight
Wall mounting type: Approx. 0.58 kg/Rack mounting type: Approx. 0.65 kg
* Specifications subject to changes without notice.
※
Initial clear : The start-up action of RM-6003T is required for approx.210 seconds.