4 How to Use
4-5. How to wire
- 23 -
RM-6000
<Grounding>
Connect the indicator/alarm unit to your grounding terminal.
Grounding terminal
Grounding terminal
WARNING
Before turning on the indicator/alarm unit, never fail to connect it to a grounding terminal.
For stable operation of the indicator/alarm unit and safety, it must be connected to a grounding
terminal. Do not connect the grounding wire to a gas pipe. The grounding must be made as D type
grounding (below 100
Ω
of grounding resistance).