Rigol DP700 Series Скачать руководство пользователя страница 8

RIGOL

 

Chapter 1 Calibration Notices

 

1-2

 

 

DP700 Calibration Guide 

To Stop the Calibration 

 

During calibration, you can stop the calibration in the following two ways: 

 

 

Exit the calibration page by pressing keys or using the knob. 

 

In the calibration page, use the Up arrow key or the knob to switch to the 

"TestCal" tab, then use the Left/Right arrow key or the knob to switch between 

tabs and exit the calibration page (For the save operation and precautions 

during calibration, refer to Chapter 2 in this manual). 

 

 

Turn off the instrument.

 

 

Press the Power key 

  to turn off the instrument. 

 

When you stop calibration, the saved calibration data will not be lost. You do not 

need to calibrate the items already calibrated again in the next calibration. 

 

 

CAUTION

 

If you stop calibration when the instrument is writing the new calibration 

factors into the internal non-volatile memory, you may lose all the 

calibration factors and you need to perform the calibration for all the 

required items again. 

 

To Acquire the Calibration Service 

 

RIGOL

 does not recommend users to perform manual calibration by themselves. If 

calibration is required, please contact 

RIGOL

 customer service department or the 

local distributor. 

 

Recommended Calibration Devices 

 

You are recommended to use the devices listed in Table 1-1 or other devices whose 

performance specifications meet the "Performance Requirement" listed in Table 1-1. 

 

Table 1-1

 Recommended Calibration Devices

 

Device 

Performance Requirement 

Recommended Model 

Digital Multimeter 

6

1

/

2

 

RIGOL

 DM3068 

Test Probe 

Connect  the output terminals of 

DP700 series and the input 

terminals of the multimeter

 

-- 

 

 

 

Содержание DP700 Series

Страница 1: ...RIGOL Calibration Guide DP700 Series Programmable Linear DC Power Supply July 2016 RIGOL TECHNOLOGIES INC...

Страница 2: ......

Страница 3: ...on is subject to change without notice RIGOL shall not be liable for either incidental or consequential losses in connection with the furnishing use or performance of this manual as well as any inform...

Страница 4: ...alibration Process This chapter gives a detailed description of the calibration page of DP series and the calibration process for each calibration item Format Conventions in this Manual The function k...

Страница 5: ...I Chapter 1 Calibration Notices 1 1 Calibration Interval 1 1 Calibration Safety 1 1 To Stop the Calibration 1 2 To Acquire the Calibration Service 1 2 Recommended Calibration Devices 1 2 Calibration P...

Страница 6: ......

Страница 7: ...en configured with a default calibration password 123456 The password is stored in the internal non volatile memory and will not be changed at power off or remote interface reset You can also modify t...

Страница 8: ...ain in the next calibration CAUTION If you stop calibration when the instrument is writing the new calibration factors into the internal non volatile memory you may lose all the calibration factors an...

Страница 9: ...rform all the calibration procedures Ensure that the ambient temperature stays between 18 C and 28 C Ideally the calibration should be performed at 23 C 1 C Ensure that the ambient relative humidity i...

Страница 10: ......

Страница 11: ...a The detailed operation method for each step is as follows Step 1 Enter the calibration page 1 Warm up DP700 series for at least 30 minutes 2 Press System to enter the system utility function interfa...

Страница 12: ...ibration items The calibration items for DP700 series are available in Table 2 1 You can select one or multiple items for calibration RIGOL recommends that you perform the calibration for all the avai...

Страница 13: ...are reference voltage current values At this time use the num key or the knob for a negative value input use the knob to enter the measurement value of the digital multimeter For setting methods refer...

Страница 14: ...the currently calibrated data will not be saved and then you exit the calibration page Figure 2 4 Save the calibration data Dialog Box Note If you want to calibrate multiple items RIGOL suggests you c...

Страница 15: ...tive terminal of the digital multimeter with the terminal of the channel output of DP711 and connect the negative terminal of the digital multimeter with the terminal of the channel output of DP711 2...

Страница 16: ...ding to the ID sequence then repeat Step 5 and 6 until you complete the calibration operation for all the calibration points 8 Press the Down arrow key or use the knob to switch the parameter focus to...

Страница 17: ...to Auto 4 Enter the DP711 calibration page and select V ADC then press the Right arrow key The cursor switches to ID of the first calibration point 5 Press OK The Meas value measurement value of the...

Страница 18: ...inal of the digital multimeter with the terminal of the channel output of DP711 and connect the negative terminal of the digital multimeter with the terminal of the channel output of DP711 2 Press On...

Страница 19: ...ding to the ID sequence then repeat Step 5 and 6 until you complete the calibration operation for all the calibration points 8 Press the Down arrow key or use the knob to switch the parameter focus to...

Страница 20: ...and the scale to A 4 Enter the DP711 calibration page and select I ADC then press the Right arrow key The cursor switches to ID of the first calibration point 5 Press OK The Meas value measurement val...

Страница 21: ...ion point there is only one calibration point for OVP calibration 3 Press OK and then the On Off key on the front panel of DP711 is illuminated automatically Then the instrument starts to perform auto...

Отзывы: