RIGOL

 

Chapter 2 Calibration Process 

2-10

 

 

DP700 Calibration Guide 

I ADC 

 

This section takes DP711 as an example to introduce the calibration method for the 

current ADC of DP700 series. Unless otherwise specified, this calibration method 

applies to all the models of the DP700 series. 

 

(1)

 

Ensure that DP711 has been warmed up for at least 30 minutes. Use the test 

probe to connect the output terminals of DP711 to the current input terminals of 

the digital multimeter, as shown in Figure 2-6. 

 

(2)

 

Press On/Off on the front panel of DP711 to turn on the channel output. 

 

(3)

 

Turn on the digital multimeter. Select the DC current (DCI) measurement 

function and set the range to "Auto" and the scale to "A". 

 

(4)

 

Enter the DP711 calibration page, and select "I ADC", then press the Right arrow 

key. The cursor switches to ID of the first calibration point. 

 
(5)

 

Press OK. The Meas value (measurement value) of the selected calibration point 

is highlighted and editable. Then, use the num key or the knob to enter the 

measurement value of the digital multimeter. 

 

(6)

 

Press OK, and the measurement value will be written into the system. Thus, you 

have completed the calibration for the selected calibration point. 

 

(7)

 

Select the rest of the calibration points according to the ID sequence, then 

repeat Step (5) and (6) until you complete the calibration operation for all the 

calibration points. 

 

(8)

 

Press the Down arrow key or use the knob to switch the parameter focus to 

"Save", then, press OK to save the current calibration data. Till now, you have 
completed the current ADC calibration. 

 

 

Содержание DP700 Series

Страница 1: ...RIGOL Calibration Guide DP700 Series Programmable Linear DC Power Supply July 2016 RIGOL TECHNOLOGIES INC...

Страница 2: ......

Страница 3: ...on is subject to change without notice RIGOL shall not be liable for either incidental or consequential losses in connection with the furnishing use or performance of this manual as well as any inform...

Страница 4: ...alibration Process This chapter gives a detailed description of the calibration page of DP series and the calibration process for each calibration item Format Conventions in this Manual The function k...

Страница 5: ...I Chapter 1 Calibration Notices 1 1 Calibration Interval 1 1 Calibration Safety 1 1 To Stop the Calibration 1 2 To Acquire the Calibration Service 1 2 Recommended Calibration Devices 1 2 Calibration P...

Страница 6: ......

Страница 7: ...en configured with a default calibration password 123456 The password is stored in the internal non volatile memory and will not be changed at power off or remote interface reset You can also modify t...

Страница 8: ...ain in the next calibration CAUTION If you stop calibration when the instrument is writing the new calibration factors into the internal non volatile memory you may lose all the calibration factors an...

Страница 9: ...rform all the calibration procedures Ensure that the ambient temperature stays between 18 C and 28 C Ideally the calibration should be performed at 23 C 1 C Ensure that the ambient relative humidity i...

Страница 10: ......

Страница 11: ...a The detailed operation method for each step is as follows Step 1 Enter the calibration page 1 Warm up DP700 series for at least 30 minutes 2 Press System to enter the system utility function interfa...

Страница 12: ...ibration items The calibration items for DP700 series are available in Table 2 1 You can select one or multiple items for calibration RIGOL recommends that you perform the calibration for all the avai...

Страница 13: ...are reference voltage current values At this time use the num key or the knob for a negative value input use the knob to enter the measurement value of the digital multimeter For setting methods refer...

Страница 14: ...the currently calibrated data will not be saved and then you exit the calibration page Figure 2 4 Save the calibration data Dialog Box Note If you want to calibrate multiple items RIGOL suggests you c...

Страница 15: ...tive terminal of the digital multimeter with the terminal of the channel output of DP711 and connect the negative terminal of the digital multimeter with the terminal of the channel output of DP711 2...

Страница 16: ...ding to the ID sequence then repeat Step 5 and 6 until you complete the calibration operation for all the calibration points 8 Press the Down arrow key or use the knob to switch the parameter focus to...

Страница 17: ...to Auto 4 Enter the DP711 calibration page and select V ADC then press the Right arrow key The cursor switches to ID of the first calibration point 5 Press OK The Meas value measurement value of the...

Страница 18: ...inal of the digital multimeter with the terminal of the channel output of DP711 and connect the negative terminal of the digital multimeter with the terminal of the channel output of DP711 2 Press On...

Страница 19: ...ding to the ID sequence then repeat Step 5 and 6 until you complete the calibration operation for all the calibration points 8 Press the Down arrow key or use the knob to switch the parameter focus to...

Страница 20: ...and the scale to A 4 Enter the DP711 calibration page and select I ADC then press the Right arrow key The cursor switches to ID of the first calibration point 5 Press OK The Meas value measurement val...

Страница 21: ...ion point there is only one calibration point for OVP calibration 3 Press OK and then the On Off key on the front panel of DP711 is illuminated automatically Then the instrument starts to perform auto...

Отзывы: