![Rigol DP700 Series Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/rigol/dp700-series/dp700-series_calibration-manual_1461092012.webp)
RIGOL
2-2
DP700 Calibration Guide
Figure 2-2
Calibration Password Input Box
(4)
Press OK to confirm your password, and then you enter the calibration
page, as shown in Figure 2-3.
Figure 2-3
Calibration Page
Step 2 Select the calibration item
In the calibration page, press Up/Down arrow key or use the knob to switch the
calibration items. The calibration items for DP700 series are available in Table
2-1. You can select one or multiple items for calibration.
RIGOL
recommends
that you perform the calibration for all the available items.
Table 2-1
Calibration Items
No.
Calibration Item
1
V DAC (calibrate the accuracy of the output voltage)
2
I DAC (calibrate the accuracy of the output current)
3
V ADC (calibrate the accuracy of the readback voltage)
4
I ADC (calibrate the accuracy of the readback current)
5
OVP (calibrate the accuracy of the overvoltage protection setting value)