DP700 Calibration Guide
Step 6 Save the calibration data
In the calibration page, press the Up/Down arrow key or use the knob to switch
the parameter focus to "Save", then, press OK to save the current calibration
data and exit the calibration page.
If you calibrated the items but did not save your calibration data, when you exit
the calibration page, a dialog box "Save the calibration data?" will be displayed,
as shown in Figure 2-4. If you select "YES", the currently calibrated data will be
saved and then you exit the calibration page; if "NO", the currently calibrated
data will not be saved and then you exit the calibration page.
Figure 2-4 "Save the calibration data?" Dialog Box
Note:
If you want to calibrate multiple items,
RIGOL
suggests you complete the
calibration for all the items and then save the calibration data, so that you
do not have to enter the calibration password repeatedly to enter the
calibration page for calibrating each calibration item.
To guarantee the calibration results, perform the save operation for the
calibration item after all the calibration points of a certain calibration item
are all calibrated.
The following section describes the calibration method for each calibration item.
CAUTION
Please strictly follow the calibration order listed below to perform the
calibration for each item.