Chapter 2 Calibration Process 

2-4

 

 

DP700 Calibration Guide 

Step 6 Save the calibration data 

 

In the calibration page, press the Up/Down arrow key or use the knob to switch 
the parameter focus to "Save", then, press OK to save the current calibration 
data and exit the calibration page. 

 

If you calibrated the items but did not save your calibration data, when you exit 

the calibration page, a dialog box "Save the calibration data?" will be displayed, 

as shown in Figure 2-4. If you select "YES", the currently calibrated data will be 

saved and then you exit the calibration page; if "NO", the currently calibrated 

data will not be saved and then you exit the calibration page. 

 

 

Figure 2-4 "Save the calibration data?" Dialog Box 

 

Note: 

 

If you want to calibrate multiple items, 

RIGOL

 suggests you complete the 

calibration for all the items and then save the calibration data, so that you 

do not have to enter the calibration password repeatedly to enter the 

calibration page for calibrating each calibration item. 

 

 

To guarantee the calibration results, perform the save operation for the 

calibration item after all the calibration points of a certain calibration item 

are all calibrated. 

 

The following section describes the calibration method for each calibration item. 

 

 

CAUTION

 

Please strictly follow the calibration order listed below to perform the 

calibration for each item. 

Содержание DP700 Series

Страница 1: ...RIGOL Calibration Guide DP700 Series Programmable Linear DC Power Supply July 2016 RIGOL TECHNOLOGIES INC...

Страница 2: ......

Страница 3: ...on is subject to change without notice RIGOL shall not be liable for either incidental or consequential losses in connection with the furnishing use or performance of this manual as well as any inform...

Страница 4: ...alibration Process This chapter gives a detailed description of the calibration page of DP series and the calibration process for each calibration item Format Conventions in this Manual The function k...

Страница 5: ...I Chapter 1 Calibration Notices 1 1 Calibration Interval 1 1 Calibration Safety 1 1 To Stop the Calibration 1 2 To Acquire the Calibration Service 1 2 Recommended Calibration Devices 1 2 Calibration P...

Страница 6: ......

Страница 7: ...en configured with a default calibration password 123456 The password is stored in the internal non volatile memory and will not be changed at power off or remote interface reset You can also modify t...

Страница 8: ...ain in the next calibration CAUTION If you stop calibration when the instrument is writing the new calibration factors into the internal non volatile memory you may lose all the calibration factors an...

Страница 9: ...rform all the calibration procedures Ensure that the ambient temperature stays between 18 C and 28 C Ideally the calibration should be performed at 23 C 1 C Ensure that the ambient relative humidity i...

Страница 10: ......

Страница 11: ...a The detailed operation method for each step is as follows Step 1 Enter the calibration page 1 Warm up DP700 series for at least 30 minutes 2 Press System to enter the system utility function interfa...

Страница 12: ...ibration items The calibration items for DP700 series are available in Table 2 1 You can select one or multiple items for calibration RIGOL recommends that you perform the calibration for all the avai...

Страница 13: ...are reference voltage current values At this time use the num key or the knob for a negative value input use the knob to enter the measurement value of the digital multimeter For setting methods refer...

Страница 14: ...the currently calibrated data will not be saved and then you exit the calibration page Figure 2 4 Save the calibration data Dialog Box Note If you want to calibrate multiple items RIGOL suggests you c...

Страница 15: ...tive terminal of the digital multimeter with the terminal of the channel output of DP711 and connect the negative terminal of the digital multimeter with the terminal of the channel output of DP711 2...

Страница 16: ...ding to the ID sequence then repeat Step 5 and 6 until you complete the calibration operation for all the calibration points 8 Press the Down arrow key or use the knob to switch the parameter focus to...

Страница 17: ...to Auto 4 Enter the DP711 calibration page and select V ADC then press the Right arrow key The cursor switches to ID of the first calibration point 5 Press OK The Meas value measurement value of the...

Страница 18: ...inal of the digital multimeter with the terminal of the channel output of DP711 and connect the negative terminal of the digital multimeter with the terminal of the channel output of DP711 2 Press On...

Страница 19: ...ding to the ID sequence then repeat Step 5 and 6 until you complete the calibration operation for all the calibration points 8 Press the Down arrow key or use the knob to switch the parameter focus to...

Страница 20: ...and the scale to A 4 Enter the DP711 calibration page and select I ADC then press the Right arrow key The cursor switches to ID of the first calibration point 5 Press OK The Meas value measurement val...

Страница 21: ...ion point there is only one calibration point for OVP calibration 3 Press OK and then the On Off key on the front panel of DP711 is illuminated automatically Then the instrument starts to perform auto...