Chapter 2 Front Panel Operations
RIGOL
DL3000 User’s Guide
2-11
Constant Resistance (CR) Mode
In CR mode, the load is regarded as a constant resistance, and changes the current
linearly proportional to the input voltage, as shown in Figure 2-8.
Figure 2-8 Constant Resistance (CR) Mode
Note: When in CR mode, please ensure that the input device works in the constant
voltage mode.
Operation Procedures:
1.
Power on the instrument
Press the Power key
on the front panel to power on the instrument.
2.
Connect the DUT and the channel input terminals of the load
Connect the DUT and the channel input terminals on the front panel of the load
as shown in Figure 2-2.
CAUTION
While making a connection, the positive polarity of the load should
be connected to the (+) terminal of the channel output, and the
negative polarity of the load to the (-) terminal of the channel
output. A misconnection with the terminals may cause damage to
the instrument or the DUT.
3.
Set channel parameters
Press CR to enter the main interface of CR mode, as shown in Figure 2-9. At the
top of the main interface, "CR" and "STATIC" are active. Of which, "STATIC" is
highlighted.
I
Load Current
V
Set Resistance
(R=V/I)
Input Voltage