Chapter 2 Front Panel Operations 

2-4 

DG1000Z User’s Guide 

To Set Frequency/Period 

 

Frequency is one of the most important parameters of basic waveforms. For different 

instrument models and waveforms, the setting ranges of frequency are different. For 

detailed information, please refer to “Frequency Characteristics” in Specifications”

The default frequency is 1kHz.   

 

The frequency displayed on the screen is the default value or the frequency 

previously set. When the instrument function is changed, if this frequency is valid 

under the new function, the instrument will still use this frequency; otherwise, the 

instrument would display prompt message and set the frequency to the frequency 

upper limit of the new function automatically.   

 

Press Freq/Period to highlight “Freq”. At this point, use the numeric keyboard to 

input the desired frequency value and then select the desired unit from the pop-up 

menu.   

 

The frequency units available are MHz, kHz, Hz, mHz and μHz.   

 

Press this softkey again to switch to period setting. At this point, “Period” is 

highlighted.   

 

The period units available are sec, msec, μsec and nsec.   

 

Users can also use the direction keys and knob to set the parameter value: use the 

direction keys to move the cursor to select the digit to be edited and then rotate the 

knob to change the number.   

 

 

Содержание DG1022Z

Страница 1: ...RIGOL User s Guide DG1000Z Series Function Arbitrary Waveform Generator Oct 2016 RIGOL TECHNOLOGIES INC...

Страница 2: ......

Страница 3: ...is subject to change without notice RIGOL shall not be liable for either incidental or consequential losses in connection with the furnishing use or performance of this manual as well as any informat...

Страница 4: ...ing any input or output terminals Connect the Probe Correctly If a probe is used do not connect the ground lead to high voltage since it has isobaric electric potential as the ground Observe All Termi...

Страница 5: ...te in an Explosive Atmosphere To avoid personal injuries or damage to the instrument never operate the instrument in an explosive atmosphere Keep Instrument Surfaces Clean and Dry To avoid dust or moi...

Страница 6: ...s of important data Safety Terms on the Product DANGER It calls attention to an operation if not correctly performed could result in injury or hazard immediately WARNING It calls attention to an opera...

Страница 7: ...geerdet Schlie enSie die Erdungsklemmen an keine hohe Spannung an Beachten Sie alle Anschl sse Zur Vermeidung von Feuer oder Stromschlag beachten Sie alle Bemerkungen und Markierungen auf dem Instrume...

Страница 8: ...ng von Personen und Sachsch den ist es unumg nglich das Ger t ausschlie lich fernab jedweder explosiven Atmosph re zu betreiben Ger teoberfl chen sauber und trocken halten Um den Einflu von Staub und...

Страница 9: ...die Sch den am Ger t hervorrufen k nnen Begriffe auf dem Produkt DANGER weist auf eine Verletzung oder Gef hrdung hin die sofort geschehen kann WARNING weist auf eine Verletzung oder Gef hrdung hin d...

Страница 10: ...onnect the instrument from all power sources 2 Clean the external surfaces of the instrument with a soft cloth dampened with mild detergent or water When cleaning the LCD take care to avoid scarifying...

Страница 11: ...The equipment may contain substances that could be harmful to the environment or human health To avoid the release of such substances into the environment and avoid harm to human health we recommend...

Страница 12: ...RIGOL X DG1000Z User s Guide...

Страница 13: ...full functional channels which are equivalent to two independent generators 1ppm frequency stability 125dBc Hz phase noise Built in 8 orders harmonics generator Built in 7 digits s full function frequ...

Страница 14: ...00Z series Chapter 6 Appendix Provide the information about the options and accessories list as well as warranty information of DG1000Z Format Conventions in this Manual 1 Button The button at the fro...

Страница 15: ...z DG1032Z 2 30MHz DG1022Z 2 25MHz 2 Both models of DG1000Z series function arbitrary waveform generator are equipped with dual channels CH1 and CH2 Unless otherwise specified this manual takes CH1 as...

Страница 16: ...and Inspection 1 14 To Connect to Power 1 14 Power on 1 14 To Set the System Language 1 14 User Interface 1 15 Dual Channels Parameters Mode 1 15 Dual Channels Graph Mode 1 18 Single Channel View Mod...

Страница 17: ...odulation FM 2 39 Phase Modulation PM 2 42 Amplitude Shift Keying ASK 2 45 Frequency Shift Keying FSK 2 48 Phase Shift Keying PSK 2 51 Pulse Width Modulation PWM 2 54 Sweep 2 57 To Enable Sweep Functi...

Страница 18: ...Set 2 100 I O Configuration 2 103 Print Set 2 107 Test Calibration 2 107 To Use External Power Amplifier Option 2 108 To Install the Option 2 112 To Lock the Keyboard 2 114 Chapter 3 Remote Control 3...

Страница 19: ...ly introduces the appearance and dimensions front rear panel and user interface of DG1000Z Subjects in this chapter General Inspection To Adjust the Handle Appearance and Dimensions Front Panel Overvi...

Страница 20: ...rier shall be liable for the damage to the instrument resulting from shipment RIGOL would not be responsible for free maintenance rework or replacement of the instrument 2 Inspect the instrument In ca...

Страница 21: ...1 3 To Adjust the Handle To adjust the handle please hold the handle by sides of the instrument and pull it outward and then rotate the handle to the desired position as shown in the figure below Adju...

Страница 22: ...RIGOL Chapter 1 Quick Start 1 4 DG1000Z User s Guide Appearance and Dimensions Front View Unit mm Side View Unit mm...

Страница 23: ...orm or state files saved in the USB storage device or store the current instrument states or edited waveform data into the USB storage device In additional the content displayed on the screen also can...

Страница 24: ...nfiguration of CH1 Press this key again to close the output of CH1 and the backlight turns off It is used to control the output of CH2 Press this key to open the output of CH2 the backlight turns on a...

Страница 25: ...sor when inputting parameter using numeric keyboard Used to unfold or fold the current selected directory when storing or reading file Used to move the cursor to select the specified character in file...

Страница 26: ...Period Ampl HiLevel Offset LoLevel Symmetry and Start Phase of ramp waveform Output Pulse with frequency from 1 Hz to 25MHz and variable pulse width and edge time The backlight turns on when this fun...

Страница 27: ...e used to generate Gated Burst 3 types of trigger sources Internal External and Manual The backlight turns on when this function is selected Used to set the auxiliary function parameters and system pa...

Страница 28: ...displayed menus respectively Press this softkey to activate the corresponding menu 16 LCD 3 5 inches TFT 320 240 color LCD display The current function menu settings system state as well as prompt me...

Страница 29: ...about the characteristics of the sync signals corresponding to various output signals refer to the introduction in Sync Set 2 Ext Mod When AM FM PM or PWM of CH1 is enabled and external modulation so...

Страница 30: ...input impedance is 1000 For the detailed introduction refer to Modulation 3 FSK When ASK FSK or PSK of CH2 is enabled and external modulation source is selected this connector accepts an external mod...

Страница 31: ...ol An integrated testing system may be built as the generator conforms to the LXI Core 2011 Device class standard of LAN based instrument control 5 USB Device Used to connect the generator to a comput...

Страница 32: ...rrect voltage range automatically and users do not need to select the voltage range manually Figure 1 3 Connect to Power CAUTION To avoid electric shock make sure that the instrument is correctly grou...

Страница 33: ...ut Configuration Status Bar Display the current output configurations of the two channels 7 8 9 10 Type of Output Impedance High impedance display HighZ Load display impedance the default is 50 and th...

Страница 34: ...connected with power amplifier correctly 5 Waveform Display the waveform currently selected in each channel 6 Channel Status Bar Used to indicate the selected status and on off status of the channels...

Страница 35: ...of the channel Press Offset LoLevel to highlight Offset and use the numeric keyboard or direction keys and knob to modify this parameter 10 Phase Display the waveform start phase of the channel Press...

Страница 36: ...lity System Display DispMod to select Dual Graph as shown in the figure below Figure 1 5 User Interface Dual Channels Graph Mode Single Channel View Mode Press Utility System Display DispMode to selec...

Страница 37: ...displayed 2 Page Up Down When the help information is displayed in multiple pages users can acquire the help information on the previous or next page using the previous line the next line page up page...

Страница 38: ......

Страница 39: ...s This chapter introduces the main functions and operation methods of DG1000Z Subjects in this chapter To Output Basic Waveform To Output Arbitrary Waveform To Output Harmonic Modulation Sweep Burst C...

Страница 40: ...used to switch the currently selected channel between CH1 and CH2 At start up CH1 is selected by default and the corresponding area in the user interface is highlighted and the border of the channel s...

Страница 41: ...eform At this point the backlight of the button turns on and the corresponding function name and parameter setting menu as shown in the table below are displayed at the right of the user interface At...

Страница 42: ...the instrument will still use this frequency otherwise the instrument would display prompt message and set the frequency to the frequency upper limit of the new function automatically Press Freq Peri...

Страница 43: ...yboard to input the desired amplitude value and then select the desired unit from the pop up menu The amplitude units available are Vpp mVpp Vrms mVrms and dBm invalid in HighZ Press this softkey agai...

Страница 44: ...and set proper load value using numeric keyboard 3 Select the desired waveform and press Ampl HiLevel to highlight Ampl Use the numeric keyboard to input the desired value and then select dBm from th...

Страница 45: ...age and set the offset to the offset upper limit of the new configuration automatically Press Offset LoLevel to highlight Offset At this point use the numeric keyboard to input the desired offset valu...

Страница 46: ...hen the instrument function is changed the new function will still use this phase Press Start Phase to highlight the softkey At this point use the numeric keyboard to input the desired start phase val...

Страница 47: ...waveform 1kHz 5Vpp 0 is outputed from CH1 while another one 1kHz 5Vpp 180 from CH2 Use an oscilloscope to sample and display the two signals you will see that the waveforms shown on the oscilloscope...

Страница 48: ...range of duty cycle is limited by the Freq Period setting Please refer to Signal Characteristics in Specifications The default value is 50 Press Duty Cycle to highlight the softkey At this point use...

Страница 49: ...selected T t Symmetry t T 100 The setting range of symmetry is from 0 to 100 and the default is 50 Press Symm to highlight the softkey At this point use the numeric keyboard to input the desired symme...

Страница 50: ...mum Pulse Width 2 Pulse duty cycle is defined as the percentage that the pulse width takes up in the whole period Pulse duty cycle and pulse width are correlative Once a parameter is changed the other...

Страница 51: ...rently specified pulse width limit as shown in the formula below DG1000Z will automatically adjust the edge time to match the specified pulse width if the value currently set exceeds the limit value L...

Страница 52: ...g waveform output you can also configure parameters such as Impedance and Polarity related to the channel output through Channel Set menu in Utility For details please refer to the introduction in Out...

Страница 53: ...eyboard to input 20 Then select kHz from the pop up menu 4 To set the amplitude Press Ampl HiLevel to highlight Ampl and use the numeric keyboard to input 2 5 Then select Vpp from the pop up menu 5 To...

Страница 54: ...RIGOL Chapter 2 Front Panel Operations 2 16 DG1000Z User s Guide Figure 2 3 Sine Waveform...

Страница 55: ...a the PC software download the waveform to the instrument via the remote interface or store the waveform in a USB storage device and then read and output the waveform via the instrument To Enable Arbi...

Страница 56: ...switch to frequency output mode automatically 2 Sample Rate Output Mode Press Arb Arb Mode to select the Sample Rate mode In sample rate output mode users can set the sample rate the points outputted...

Страница 57: ...form the operations below 1 To output the selected waveform Press the corresponding output control key Output1 or Output2 and the specified waveform will be outputted from the channel 2 To edit the se...

Страница 58: ...nd rotate the knob to select the desired waveform the selected waveform is highlighted Then press Select to select the waveform which is highlighted Figure 2 5 Built in Waveform Table 2 2 160 Built in...

Страница 59: ...terworth filter Chebyshev1 Chebyshev1 filter Chebyshev2 Chebyshev2 filter Signal TV TV signal Voice Voice signal Surge Surge signal Radar Analog radar waveform DualTone Dual tone signal Ripple Ripple...

Страница 60: ...3 Waveform 3 of the nerve stimulation electrotherapy AutoElec Auto Ignition Ignition waveform of the automotive motor ISO16750 2 SP Automotive starting profile with ringing ISO16750 2 VR Automotive su...

Страница 61: ...eigh distribution Cauchy Cauchy distribution function TF CosH Hyperbolic cosine CosInt Integral cosine Cot Cotangent CotHCon Concave hyperbolic cotangent CotHPro Protuberant hyperbolic cotangent CscCo...

Страница 62: ...hev window FlattopWin Flat Top weighted window Hamming Hamming window Hanning Hanning window Kaiser Kaiser window NuttallWin Nuttall defined minimum 4 term Blackman Harris window ParzenWin Parzen wind...

Страница 63: ...up the volatile memory with the waveform currently edited through Edit Wform 2 Press Arb Edit Wform Data Src to select CH1 Arb CH2 Arb or Vol Then copy the currently selected arbitrary waveform of CH...

Страница 64: ...al number of points of the waveform currently edited is always 8192 The period range is from 50ns to 1Ms and the default is 1ms For the setting method refer to To Set Frequency Period 2 High Level It...

Страница 65: ...to the 8192nd point Note When the number of points is modified the original waveform data in the volatile memory will be lost 5 Data Source Press Data Src to select CH1 Arb CH2 Arb or Volatile You ca...

Страница 66: ...he Edit Points interface Edit Points Select the number 1 to Sa of the point to be edited and input the desired value using numeric keyboard or direction keys and knob Voltage Set the voltage of the cu...

Страница 67: ...e voltage the units available are mV and V of the start point of the block The range is limited by the current High Level and Low Level X2 Set the number of the end point of the block X2 should be les...

Страница 68: ...waveform phase The frequencies of the other components called harmonics are all integral multiples of the fundamental waveform frequency Components whose frequencies are odd multiples of the fundament...

Страница 69: ...d harmonic order The range is limited by the maximum output frequency of the instrument and current fundamental waveform frequency Range 2 to maximum output frequency of the instrument current fundame...

Страница 70: ...pecified To Set Harmonic Amplitude After entering the harmonic setting menu press Harmonic Ampl to set the amplitude of each order of harmonic 1 SN press this softkey to select the sequence number of...

Страница 71: ...the harmonic setting menu 4 To Set Harmonic Order In the harmonic setting menu press Order and use the numeric keyboard to input 5 before pressing OK 5 To Select Harmonic Type In the harmonic setting...

Страница 72: ...fundamental waveform 2th and 4th order of harmonics will be outputted from the CH1 connector according to the current configuration 9 To Observe the Output Waveform Connet the CH1 of the DG1000Z to a...

Страница 73: ...rm Shape AM carrier waveform could be Sine Square Ramp or Arbitrary waveform except DC The default is Sine Press the front panel key Sine Square Ramp or Arb you need to press Select Wform in the arbit...

Страница 74: ...lt is Sine Square 50 duty cycle Triangle 50 symmetry UpRamp 100 symmetry DnRamp 0 symmetry Arb the arbitrary waveform selected of the current channel Note noise can be used as modulating waveform but...

Страница 75: ...c keyboard or direction keys and knob The modulating waveform frequency ranges from 2mHz to 1MHz and the default value is 100Hz Note when external modulation source is selected this menu will be graye...

Страница 76: ...waveform components carry no information the modulation is less efficient In order to improve the modulation efficiency the carrier waveform components are suppressed on the base of the normal amplit...

Страница 77: ...sed as carrier waveform To Set Carrier Waveform Parameters The different settings of various parameters frequency amplitude DC offset start phase and etc of the carrier waveform will influence the FM...

Страница 78: ...ry waveform selected of the current channel Note noise can be used as modulating waveform but can not be used as carrier waveform 2 External Source When external modulation source is selected Shape an...

Страница 79: ...r than or equal to the carrier frequency The sum of frequency deviation and carrier frequency must be lower than or equal to the sum of the current carrier frequency upper limit and 1kHz Note if Sine...

Страница 80: ...ier Waveform Parameters The different settings of various parameters frequency amplitude DC offset etc of the carrier waveform will influence the PM modulated waveform For different carrier waveform s...

Страница 81: ...itrary waveform selected of the current channel Note noise can be used as modulating waveform but can not be used as carrier waveform 2 External Source When external modulation source is selected Shap...

Страница 82: ...ed To Set Phase Deviation Phase deviation is the deviation of the modulating waveform phase relative to the carrier waveform phase Press PM Dev to set the PM phase deviation Use the numeric keyboard o...

Страница 83: ...shape Pulse Noise and DC could not be used as carrier waveform To Set Carrier Waveform Parameters The different settings of various parameters frequency amplitude DC offset and etc of the carrier wave...

Страница 84: ...t Mod Trig FSK connector at the rear panel Note for the CH1 Sync Ext Mod Trig FSK connector controlling ASK modulation externally is different from controlling AM FM PM modulations externally In ASK m...

Страница 85: ...nternal modulation set the polarity to Pos and the generator would output the lower of the carrier amplitude and modulating amplitude when the modulating waveform is logic low level and output the gre...

Страница 86: ...e Noise and DC could not be used as carrier waveform To Set Carrier Waveform Parameters The different settings of various parameters frequency amplitude DC offset start phase and etc of the carrier wa...

Страница 87: ...od Trig FSK connector at the rear panel Note for the CH1 Sync Ext Mod Trig FSK connector controlling FSK modulation externally is different from controlling AM FM PM modulations externally In FSK modu...

Страница 88: ...Pos or Neg polarity of the modulating waveform to control the output frequency In internal modulation set the polarity to Pos and the generator would output the carrier frequency when the modulating...

Страница 89: ...e and DC could not be used as carrier waveform To Set Carrier Waveform Parameters The different settings of various parameters frequency amplitude DC offset start phase and etc of the carrier waveform...

Страница 90: ...xt Mod Trig FSK connector at the rear panel Note for the CH1 Sync Ext Mod Trig FSK connector controlling PSK modulation externally is different from controlling AM FM PM modulations externally In PSK...

Страница 91: ...utput phase In internal modulation set the polarity to Pos and the generator would output the carrier phase when the modulating waveform is logic low level and output the modulating phase when the mod...

Страница 92: ...ll select PWM automatically When Mod is enabled Sweep or Burst will be automatically disabled if enabled currently To Select Carrier Waveform Shape As mentioned before PWM can only be used to modulate...

Страница 93: ...will be grayed out and disabled The generator accepts the external modulating signal from the CH1 Sync Ext Mod Trig FSK connector at the rear panel At this time the width deviation or duty cycle devia...

Страница 94: ...viation is limited by the minimum pulse width and current edge time setting Duty cycle deviation represents the variation in of the modulated waveform duty cycle relative to the original pulse duty cy...

Страница 95: ...text for details Start Frequency and Stop Frequency Start frequency and stop frequency are the upper and lower limits of the frequency for frequency sweep The generator always sweeps from the start f...

Страница 96: ...m frequency of the waveform currently selected as min F the maximum frequency as max F and 2 max min F F Fm The range of center frequency is from min F to max F and the parameters for different wavefo...

Страница 97: ...arithmic fashion that is the output frequency changes in the way of octave per second or decade per second The variation is controlled by Start Frequency Stop Frequency and Sweep Time When Log Sweep i...

Страница 98: ...2 and the range is from 2 to 1024 Note in Linear and Log sweep modes Step Num is grayed out and disabled Figure 2 9 Step Sweep Sweep Time When Sweep is enabled press SweepTime and use the numeric keyb...

Страница 99: ...Start Frequency once the mark frequency is modified KEY POINTS For step sweep the sweep points determined by the start frequency stop frequency and step number respectively are f1 f2 fn fn 1 if the m...

Страница 100: ...trigger signal is received and then wait for the next trigger source When Sweep is enabled press Trigger Source to select Int Ext or Manual The default is Int 1 Internal Trigger The generator outputs...

Страница 101: ...ges from low level to high level at the start of each sweep and returns to low level at center frequency point or at the specified mark frequency point 2 When Leading is selected the CH1 Sync Ext Mod...

Страница 102: ...select Leading or Trailing The CH1 Sync Ext Mod Trig FSK connector at the rear panel is used as the input terminal of external trigger signal The instrument generates a sweep at the leading edge of t...

Страница 103: ...lease refer to the introduction in the following text for details Burst Type DG1000Z can output three types of bursts including N cycle infinite and gated The default is N cycle Table 2 3 Relations am...

Страница 104: ...ect Infinite and the instrument will set the trigger source to Manual automatically An infinite cycle burst sketch will appear on the screen Figure 2 11 Infinite Burst For infinite burst Ext or Manual...

Страница 105: ...Here waveform period is the period of burst function such as Sine and Square If the burst period currently set is too short the generator will increase this period automatically to allow the output o...

Страница 106: ...trigger When Burst is enabled press Trigger Source to select Int Ext or Manual The default is Int 1 Internal Trigger When internal trigger is selected the generator can only output N cycle burst and...

Страница 107: ...shown in the figure below The level is defined as idle level Figure 2 13 Definition of Idle Level When Burst is enabled press Idle Level to select your desired menu to set the idle level The default...

Страница 108: ...addition the dual channels can still output normally when the counter is enabled To Enable the Counter Press Counter at the front panel the backlight turns on and the left indicator flickers to enabl...

Страница 109: ...ter can measure the following parameters frequency period duty cycle positive pulse width and negative pulse width The default is frequency 3 Statistic Function Press Statist to enable or disable stat...

Страница 110: ...Display Mode Note when the statistic function is disabled Display is grayed out and disabled Figure 2 16 Statistic Results Interface In Curve Display Mode 2 To clear the statistic results Press Clear...

Страница 111: ...in low frequency signal measurement Press HFSuppre to enable or disable the high frequency suppression function Note enable high frequency suppression when low frequency signal with lower than 250kHz...

Страница 112: ...10 storage locations for state files and 10 storage locations for arbitrary waveform files Users can store instrument states and arbitrary waveform files created by users or downloaded via remote comm...

Страница 113: ...rbitrary waveform in internal or external memory in RAF format At most 10 arbitrary files can be stored in internal memory In arbitrary waveform file the voltage values corresponding to each waveform...

Страница 114: ...The instrument will enter the Arb interface automatically when the Read operation is finished The arbitrary waveform is stored in volatile memory Note only when D Disk is selected this menu is availab...

Страница 115: ...When File is selected as the browser type users can perform a series of operations on files including Save Read Copy Paste Delete and New Folder Save 1 To Select File Type Follow the introduction in F...

Страница 116: ...Chinese Input Press on the numeric keyboard at the front panel to switch to lowercase Figure 2 19 Filename Input Area Chinese Use the knob to select the desired character from the virtual soft keyboa...

Страница 117: ...se the knob to select the directory in which the file to be read is stored Then set Browser to File and use the knob to select the file to be read 3 To Read File Press Read the generator will read the...

Страница 118: ...ly selected file or folder empty folder New Folder DG1000Z allows users to create new folders in external memory Please first insert a USB storage device and make sure that it is identified by the ins...

Страница 119: ...G1000Z as Dir and use the knob of DG1000Z to move the focus cursor onto the model identifier of DS1 Now all channels and their on off status of DS1 will be displayed in the File Folder Display Area Se...

Страница 120: ...s stored in the internal volatile memory of the current channel of DG1000Z 4 Connect the output of the current channel of DG1000Z with the input of DS2 Turn on the channel output of DG1000Z and DG1000...

Страница 121: ...Set Set the output parameters of CH1 and CH2 Coupling Set Set the channel coupling parameters Channel Copy Set the channel copy parameters Set To Default Restore the instrument to its default state La...

Страница 122: ...e output of the sync signal The default is On namely transmitting the sync signal to the CH1 Sync Ext Mod Trig FSK connector When sync signal is disabled the output level at the CH1 Sync Ext Mod Trig...

Страница 123: ...onding value of the output signal is greater than the reference When the frequency of the basic waveform is less than or equal to 30MHz the sync signal has the same frequency as the basic waveform Whe...

Страница 124: ...o low level at the mid point of the sweep The frequency of the sync signal is the reciprocal of the sum of the specified sweep time return time start hold and end hold times If Mark is enabled For lin...

Страница 125: ...al trigger source is selected The terminal is used to accept the external trigger signal and there is no sync signal output Output Set 1 Output Polarity Set the output signal on the CH1 connector as n...

Страница 126: ...e to half of the previous value 2 5Vpp Notice that only the displayed values change with the parameter and the real output from the generator does not change 3 Output Mode Set the output mode of the C...

Страница 127: ...urce to select the desired waveform to be summed on the current waveform 3 Sum Frequency Set the frequency of waveform to be summed on the current waveform Press Utility Channel Set Sum Wforms Sum Fre...

Страница 128: ...2 The parameter relations are as shown below FCH2 FCH1 FDev FCH1 FCH2 FDev Ratio the frequency ratio of CH1 and CH2 The parameter relations are as shown below FCH2 FCH1 FRatio FCH1 FCH2 FRatio Note pl...

Страница 129: ...and you can not set the phase deviation or phase ratio 2 To Enable Coupling Function Press PhaseCpl to turn phase coupling On or Off The default is Off When the corresponding coupling functions are en...

Страница 130: ...quency amplitude or phase of CH2 minus or is divided by the deviation or ratio set is below the lower limit of frequency amplitude or phase of CH1 the generator will adjust the frequency phase or ampl...

Страница 131: ...ally At this point CH2 outputs the inverted signal of the output signal of CH1 if currently enabled Off disable the track function This is the default state Note when track function is enabled couplin...

Страница 132: ...tory Default Values Parameters Defaults Channel Parameters Current Carrier Waveform Sine Output Impedance HighZ Output Load 50 Sync Output On Sync Polarity Pos Sync Delay 0s Channel Output Normal Outp...

Страница 133: ...dit Points Points 1 Voltage 2 5V Edit Block X1 1 Y1 2 5V X2 8 Y2 2 5V Modulation Modulation Type AM AM Modulation Modulation Source Int Modulating Waveform Shape Sine Modulation Frequency 100Hz Modula...

Страница 134: ...ation Source Int Modulating Waveform Shape Sine Modulation Frequency 100Hz Width Deviation 200 s Duty Cycle Deviation 20 Sweep Sweep Type Linear Sweep Time 1s Return Time 0s Start Frequency 100Hz Stop...

Страница 135: ...n Ratio Ratio Default Channel CH1 Counter Measurement Parameter Frequency Gate Time 1ms Statistic Function Off Display Mode Digital Trigger Sensitivity 25 Trigger Level 0V Coupling Mode AC High freque...

Страница 136: ...Clock Source Int Decimal Point Dot Thousand Separator Comma Beeper On Screen Saver On Brightness 50 Contrast 25 Display Mode Dual Channels Parameters Language Factory Delivery Setting I O Configurati...

Страница 137: ...t method Press Utility Language to select the desired language This setting is stored in non volatile memory and will not be influenced by the Set To Default operation System Information Press Utility...

Страница 138: ...Out In connector at the rear panel If no valid clock source is detected the prompt message Not detect a valid external clock would be displayed and the clock source would be switched to Int You can sy...

Страница 139: ...t or comma The default is dot Thousands separator could be set as Space or None when decimal point is set as could be set as Space or None when decimal point is set as Number display format has 6 comb...

Страница 140: ...the same time in graph format Single View the various parameters and waveform of the currently selected channel are displayed in number and graph formats 4 Customize GUI DG1000Z allows users to defin...

Страница 141: ...GPIB address or LAN interface parameters or select the type of the device connected to USB interface The interface selection is stored in the non volatile memory To Set GPIB Address Each device connec...

Страница 142: ...nicate with instruments regardless of the types of the instrument interfaces GPIB USB LAN Ehternet or RS232 The GPIB USB LAN Ethernet or RS232 instrument which wants to communicate with VISA is called...

Страница 143: ...d or knob to enter your desired IP address The setting is stored in non volatile memory and will be loaded automatically when the generator is powered on the next time if DHCP and AutoIP are set as Of...

Страница 144: ...It is recommended that users acquire an available DNS address from their network administrator Press DNS Service and use the direction keys and numeric keyboard or knob to enter your desired DNS addr...

Страница 145: ...iguration During the store process the print progress bar is diaplayed on the screen Test Calibration Press Utility Test Cal to view calibration state and calibration time Besides press Utility Test C...

Страница 146: ...e connected with all DG series products to quickly build test platform and can also be used as an independent power amplifier to be used with other signal generators Main Features of PA1011 Flexible a...

Страница 147: ...instrument Zi is 50k and the range of voltage is from 10V to 10V or form 1 25V to 1 25V separately while the voltage gain is set to X1 or X10 Inputs that exceed these ranges may cause damages to the...

Страница 148: ...sure the vents at both sides and the fan aperture at the rear panel are visible in operation for normal working PA1011 Accessories Name Qty Description Power Cord 1 Connect the AC into the power adap...

Страница 149: ...thout gain and X10 means to amplify the signal 10 times and then output 3 Output Select Normal or Invert output mode for the signal output from the PA Pay attention to distinguish it with Output Polar...

Страница 150: ...mber of the instrument and verification code and click Generate to acquire the option license 2 Install the option You can install the option via the following 2 methods 1 Install the option by readin...

Страница 151: ...Recall Select D Disk set the file type to All select the installation file metioned above press Read 2 Install the option by sending SCPI commands Open the remote control window and send the following...

Страница 152: ...M9 DOT SIGN Numeric keyboard LEFT RIGHT KNOB Direction keys and knob OUTPUt1 OUTPUt2 CH Output control keys and channel switch key RETURN Return to the previous menu key COUNTER Counter key ALL All th...

Страница 153: ...Edit the command file according to the above requirements and store it into a USB storage device Turn on the instrument and insert the USB storage device Press Store to enter the store and recall int...

Страница 154: ......

Страница 155: ...Sigma of RIGOL Measurement Automation Explorer of NI National Instruments Corporation or Agilent IO Libraries Suite of Agilent Agilent Technologies Inc to send commands to control the instrument remot...

Страница 156: ...is a USB TMC device Assuming that your PC has already been installed with Ultra Sigma after you connect the generator to the PC and turn both on for the first time the generator is automatically confi...

Страница 157: ...Chapter 3 Remote Control RIGOL DG1000Z User s Guide 3 3 7 5 6...

Страница 158: ...es 4 View the device resource The resources found will appear under the RIGOL Online Resource directory and the model number and USB interface information of the instrument will also be displayed as s...

Страница 159: ...cally 3 Manual IP mode Enable the Manual IP mode and disable the DHCP mode as well as the Auto IP mode to set the network parameters manually If the generator is connected to the PC directly set the I...

Страница 160: ...e Start up the Ultra Sigma and click The window as shown in figure a is displayed Click and the software searches for the instrument resources currently connected to the LAN and the resources found ar...

Страница 161: ...to turn on the remote command control panel through which you can send commands and read data 6 Load LXI webpage As this generator conforms to LXI Core 2011 Device standards you can load LXI webpage t...

Страница 162: ...the GPIB terminal of it to the GPIB card terminal of the PC 2 Install the driver of GPIB card Install the driver of the GPIB card which has been connected to the PC correctly 3 Set the GPIB address P...

Страница 163: ...munication works normally if not please follow the corresponding prompt messages to solve the problem 5 View device resource Click to return back to the main interface of Ultra Sigma The resources fou...

Страница 164: ......

Страница 165: ...ic keyboard or direction keys and knob respectively 3 The generator is locked 1 Check whether the generator is working in remote mode in remote mode is displayed in the status bar of the user interfac...

Страница 166: ...ot passed 1 Check whether the generator is within calibration period 1 year 2 Check whether the generator is warmed up for at least 30 minutes before the test 3 Check whether the generator is under th...

Страница 167: ...luding Sinc Exponential Rise Exponential Fall ECG Gauss HaverSine Lorentz Dual Tone etc Frequency Characteristics Sine 1 Hz to 25MHz 1 Hz to 30MHz 1 Hz to 60MHz Square 1 Hz to 25MHz 1 Hz to 25MHz 1 Hz...

Страница 168: ...setting Duty Cycle 0 001 to 99 999 limited by the current frequency setting Leading Trailing Edge Time 10ns limited by the current frequency and pulse width settings Overshoot Typical 1Vpp 5 Jitter r...

Страница 169: ...overload occurs Modulation Characteristics Modulation Type AM FM PM ASK FSK PSK PWM AM Carrier Waveform Sine Square Ramp Arb except DC Source Internal External Modulating Waveform Sine Square Ramp Noi...

Страница 170: ...75mVRMS to 5Vac dc Input Bandwidth 50kHz Input Impedance 10k Burst Characteristics Carrier Waveform Sine Square Ramp Pulse Noise Arb except DC Carrier Frequency 2mHz to 25MHz 2mHz to 30MHz 2mHz to 60M...

Страница 171: ...itude Range 1 Hz to 25MHz 50mVRMS to 2 5Vac dc DC Coupling Pulse Width Minimum 20ns Resolution 5ns Duty Cycle Range Display 0 to 100 Input Characteristics Input Signal Range Breakdown Voltage 7Vac dc...

Страница 172: ...inal value Overvoltage Protection Overvoltage protection will take effect once any of the following two conditions is met The amplitude setting in the generator is greater than 2Vpp or the output offs...

Страница 173: ...USB Device LAN IP Protection IP2X Calibration Interval Recommend calibration interval is one year Authentication Information EMC In line with EN61326 1 2006 IEC 61000 3 2 2000 4 0kV Contact Discharge...

Страница 174: ......

Страница 175: ...DG1062Z Standard Accessories Power Cord USB Cable CB USBA USBB FF 150 BNC Cable CB BNC BNC MM 100 Quick Guide Options 16M Internal Memory Arb16M DG1000Z Rack Mount Kit for single instrument RM 1 DG100...

Страница 176: ...Amplifier Specifications Running Mode Constant Voltage Gain Switch between 10V 1V and 10V 10V DC Gain Error 5 Polarity Switch Normal Invert Effective Value of Sine Output Power RL 7 5 10W typical valu...

Страница 177: ...that the amplifier can output Note 2 Definition of Slew Rate input a large step function signal into the amplifier the output slope of the signal is saturated to a constant at a certain position This...

Страница 178: ...or repair of any defective products within a reasonable period of time To get repair service please contact with your nearest RIGOL sales or service office There is no other warranty expressed or imp...

Страница 179: ...Power Amplifier 2 108 Factory Default Values 2 94 Filter 2 21 Frequency 2 4 Frequency Coupling 2 90 Frequency Deviation 2 41 Frequency Modulation 2 39 Frequency Output Mode 2 18 Frequency Shift Keying...

Страница 180: ...equency 2 57 Stop Hold 2 62 Stored Waveform 2 24 Subnet Mask 2 105 Sweep 2 57 Sweep Time 2 60 Sweep Trigger Source 2 62 Symmetry 2 11 Sync Output 2 84 Test Calibration 2 107 TF 2 23 To Copy File 2 79...

Отзывы: