Auto Measurement Setting
Press Utility
Count
Auto, the instrument will set the measurement parameters:
Coupling mode, Sensitivity, Trig Level and High Frequency Restrain “On” or “Off”.
Manual Measurement Setting
Press
Utility
Count
Setup, enter the manual measurement mode.
1.
To Set the Coupling Mode
Press
Count
Setup
DC/AC
, set the coupling mode.
As shown in the following figure, the soft key with “DC” or “AC” is the current
coupling mode, each time you press the key, the coupling mode will change. For
example, the current mode is “AC”, press the key, then the mode will be “DC”.
Figure 2-63
To set the coupling mode
2.
To Set the Sensitivity
The sensitivity has three kinds: High, Medium and Low. Press Utility
Count
Setup
Sens to set the sensitivity. Press the corresponding soft key to switch
the Sensitivity.
Figure 2-64
To set the sensitivity
Suggestion:
For low amplitude signal, the “medium” or “high” sensitivity should be used.
For low frequency signal with high amplitude and slower rising edge, low
sensitivity is a better choice.