Wheel diameter measuring gauge, IDK 

 

 

IDK  [Revision 2.2]  4th September  2017 

11.  Service operation modes 

This  section  contains  the  description  of  the modes for  the  check of  device  effi-

ciency  and  calibration.  As  erroneous  actions  in  this mode  can  lead  to  invalid measure-
ment results, only specially trained personnel should perform such operations. 

11.1.  Calibration conditions 

Calibration of the  device is not necessary  in the current work.  It is neces-

sary only after producing, repairing and also after checking with negative result. 

To perform calibration the following means are necessary: 

 

Calibration  plate  with  the  deviation  from  flatness  less  then  ±0,5  um  for  250 
mm length (or Flat block RF510.11.000, figure 2); 

  Johansson gauge; 

  reference  wheel  of  the  known  diameter  (or  Reference  wheel  block 

RF510.11.ХХХХ, figure 3) 

11.2.  Get into operational modes 

  To get into the operational modes it is necessary to turn-off the device (press 

Red

 button more then 3 seconds). 

  Keeping 

Green

 button in pushed position turn-on the device (press 

Red

 but-

ton).  

 

Display shows “

CLbr.0

” message (the mode of calibration of the sensor's "

0

"). 

  To get into this mode it is necessary to press 

Green

 button (see. p. 11.2). 

  For transition to next mode it is necessary to press 

Blue

 button.  

 

Display shows “

CLbr.b

” message (the mode of device base calibration). 

  To get into this mode press 

Green

 button (see. p.11.3). 

  To get out the mode of calibration press 

Red

 button. 

11.3.  Calibration of the sensor zero 

  Zero calibration mode being enabled, the display shows the Johansson gauge 

value used for cali

bration in increments of 0.5 µm.  

  If editing of the Johansson gauge value is not required, go to the next step. To 

edit the Johansson gauge value, press the 

Blue

 button, and the digit to be ed-

ited starts blinking. Changing over between the digits is made by pressing the 

Blue

 button while changing of values is made by pressing the 

Green

 button. 

When editing is finished, press the 

Red

 button and confirm or cancel saving of 

the parameter by pressing the 

Green

 button or 

Red

 button, respectively. 

  Press  the 

Green

  button,  and  the  Johansson  gauge  value  starts  blinking, 

which means that the device must be placed onto a flat plate and a  Johans-
son gauge with nominal value of the previous step must be placed under the 
measuring tip. The Johansson gauge must be tightly fitted to the flat plate and 
supports and balls of the device must be firmly forced against the plate. 

  Press the 

Green

 button, and the display shows current reading of the sensor 

in its own coordinate system. By moving the device, assure that repeatability 
of measurement results is obtained. 

Содержание IDK Series

Страница 1: ......

Страница 2: ...he gauge 5 9 1 Gauge turn on 5 9 2 Single measurement 6 9 3 Measurement with averaging 6 9 4 Gauge turn out 7 10 Indication parameters setup 7 10 1 Image rotation 7 10 2 Image brightness set up 8 10 3...

Страница 3: ...in any other way in this case 2 Electromagnetic compatibility The gauge has been developed for use in industry and meets the requirements of the following standards EN 55022 2006 Information Technolog...

Страница 4: ...6 Complete set to be supplied Name Quantity Weight kg Wheel diameter measuring gauge IDK series 1 piece 0 5 Charger 1 piece 0 2 Manual 1 piece Case 1 piece Calibration tools option on request Flat bl...

Страница 5: ...of the segment chord the distance be tween the ball bearings centers which is obtained at placing the gauge onto the wheel The suggita of the segment is measured by means of the displacement converte...

Страница 6: ...ition 9 3 Measurement with averaging The program of the wheel diameter calculation contains an averaging algorithm that allows eliminating the surface defects influence on the diameter measuring resul...

Страница 7: ...utton pressings for 4 more minutes the gauge is turned out completely You can turn the gauge off by long pressing Red button more then 3 sec 10 Indication parameters setup 10 1 Image rotation To rotat...

Страница 8: ...ing and Red to cancel saving of the changed parameters At brightness choosing one should take into account that increased brightness enhances power consumption and decreases the period till battery re...

Страница 9: ...t mode it is necessary to press Blue button Display shows CLbr b message the mode of device base calibration To get into this mode press Green button see p 11 3 To get out the mode of calibration pres...

Страница 10: ...ade by pressing the Blue button while changing of values is made by pressing the Green button When editing is fin ished press the Red button and confirm or cancel saving of the parameter by pressing t...

Страница 11: ...assurance for the Wheel diameter gauge 24 months from the date of putting in operation warranty shelf life 12 months 14 Distributors AUSTRALIA Applied Measurement Australia Pty Ltd RAILWAY INSTRUMENTS...

Страница 12: ...Y RAILWAY INSTRUMENTS ONLY Juvan teollisuuskatu 28 FI 02920 ESPOO Finland Tel 358 400 422 900 Fax 358 9 2511 5510 steelwheel steelwheel fi www teraspyora fi FRANCE DB Innovation ALTHERIS France 26 ave...

Страница 13: ...20 PYSKOWICE Poland Tel 48 32 230 45 70 Fax 48 32 332 70 14 biuro ascorail pl www ascorail pl PORTUGAL UltraSens Qt da Portela Lt 22 1 Ap 152 3030 502 Coimbra Portugal Phone 351 239 796 277 Fax 351 23...

Страница 14: ...39 railway_marketing iem net www iem net USA CANADA MEXICO Acuity Products of Schmitt Industries Inc 2765 NW Nicolai Street Portland OR 97210 USA Tel 1 503 227 7908 Fax 1 503 223 1258 sales acuitylase...

Страница 15: ...y metro and tram wheels in the course of checkup examination repair and formation of wheel sets Measurements are made directly on rolling stock without wheel set roll out Disc brakes profile gauge IKD...

Отзывы: