Wheel diameter measuring gauge, IDK
IDK [Revision 2.2] 4th September 2017
10
RF510.11.000
Figure 2
If readings of the sensor are sufficiently stable, press the
Blue
button. Sensor
zero position is calculated in the device coordinate system, and prompt ap-
pears to save calibration results. Press the
Green
or
Red
button to confirm or
cancel saving of the results, respectively.
11.4. Calibration of the device base
Device base calibration mode being enabled, the display shows the value of refer-
ence diameter used for calibration.
If editing of the diameter value is not required, go to the next step. To edit the di-
ameter value, press the
Blue
button, and the digit to be edited starts blinking.
Changing over between the digits is made by pressing the
Blue
button while
changing of values is made by pressing the
Green
button. When editing is fin-
ished, press the
Red
button and confirm or cancel saving of the parameter by
pressing the
Green
button or
Red
button, respectively.
Press the
Green
button, and the reference diameter value starts blinking, which
means that the device must be placed onto a gage with the diameter value set at
the previous step. The device supports must be firmly forced against the reference
block.
Press the
Green
button, and the display shows current reading of the sensor in
the device coordinate system. By moving the device, assure that repeatability of
measurement results is obtained.
If readings of the instrument are sufficiently stable, press the
Blue
button. Device
base value is calculated, and prompt appears to save calibration results. Press the
Green
or
Red
button to confirm or cancel saving of the results, respectively.