43
a) Field strengths of fixed transmitters, e.g. base stations for radio (cellular/wireless)
telephones and land mobile radios, amateur radio, AM and FM broadcast and television
transmissions, cannot be theoretically predicted with accuracy. To evaluate the electro-
magnetic environment based on fixed RF transmitters, an electromagnetic assessment
should be considered. If the measured field strength at the location where the diagnostic
instruments uni® I, II, III / econom® is used exceeds the abovementioned RF compli-
ance level, the diagnostic instruments uni® I, II, III / econom® with accessories should
be observed to verify normal operation. If abnormal performance is observed, additional
measures may be required, such as reorienting or moving the diagnostic instruments
uni® I, II, III / econom®.
b) Over the frequency range from 150 kHz to 80 MHz, field strengths should be less
than 3 V/m.
Recommended distances between portable and mobile RF communications equip-
ment and the ri-scope® L.
The diagnostic instruments uni® I, II, III / econom® with accessories are intended for
use in an electromagnetic environment in which RF interference radiation is controlled.
The customer or user of the diagnostic instruments uni® I, II, III / econom® can help
to avoid electromagnetic interference by observing the minimum distance between
portable and mobile RF communications equipment (transmitters) and the diagnostic
instruments uni® I, II, III / econom® in accordance with the maximum output power of
the communication equipment.
Rated maximum
output power of the
transmitter
(W)
Separation distance according to the frequency of the trans-
mitter
(m)
150 kHz - 80
MHz
80 MHz - 800
MHz
800 MHz - 2,7 GHz
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters with a maximum output power not listed above, the recommended dis-
tance d in metres (m) can be estimated using the equation for the transmitter frequency,
where P is the maximum output power of the transmitter in watts (W). according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance applies to the higher frequency
range.
NOTE 2 These guidelines may not apply in all situations. The electromagnetic propaga-
tion is affected by absorption and reflection by structures, objects and people.
Note 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. The electromagnetic propaga-
tion is affected by absorption and reflection by structures, objects and people.
Содержание uni econom I
Страница 23: ...23...
Страница 45: ...45...
Страница 113: ...113 1 1 B MD II C F 1 2 3 4 5 6 7 8 9 1 0 11 12 2021...
Страница 114: ...114 1 3 1 4 Riester May May Riester Riester Riester 1 4 1 meatus acusticus exter nus...
Страница 115: ...115 May Riester Riester 1 4 2 1 4 3 1 4 4 1 4 5 1 4 6...
Страница 116: ...116 1 5 24 Riester 24 5 DIN EN 62471...
Страница 122: ...122 3 3 3 2 3 3 C 3 5 230 120 c 230 120 c 24 24 3 2 2 3 2 3 3 3 4 3 3 5 0 0 3 3 6...
Страница 123: ...123 3 4 1 2 3 4 5 3 4 1 3 4 2 6 3 4 3 360 4 3 4 4 3 4 5 3 4 1 3 4 1 41 C 105 8 F 3 5 1 2...
Страница 124: ...124 3 5 1 0 20 0 20 3 5 2 3 5 3 24 5 DIN EN 62471 41 C 3 6 1 3 6 1...
Страница 125: ...125 41 C 3 7 3 7 1 1 3 4 1 41 C 105 8 F 3 8 3 8 1 1 41 C 105 8 F 3 9 3 9 1 1 2 41 C 105 8 F 4 4 1...
Страница 129: ...9 EN60601 1 30 12 uni I II III econom...
Страница 132: ...IEC 61000 4 8 30 A 50 60 30 A 50 60...
Страница 136: ...10 2 RIESTER Rudolf Riester GmbH Dept Rudolf Riester Bruckstr 31 D 72417 Jungingen Germany...
Страница 137: ......
Страница 138: ......
Страница 139: ......