background image

1

Thank you for your purchase of a Richter Optica microscope.  The information in this manual is provided to answer most 

questions that can arise when operating your microscope and to help you avoid unneccesary maintenance expenses in 

the future.

Please carefully read instructions before operating microscope. Nomenclature used to describe components and controls 

are identified on opposite page.

UNPACKING

Do not discard styrofoam container or packing materials until you are sure shipment is complete and undamaged (retain 

styrofoam shipping container to store your microscope when it is not in use).  Remove all tape and packing material used 

to protect microscope during shipment.  Make certain lens surfaces do not come in contact with dirt, fingerprints or oil.  

Damage of lens surfaces occur when they come in contact with such contaminants, and image quality is reduced.

OPERATION

 

A. 

Illumination

 

 

1.

 

Before operating microscope, adjust intensity control located on side of base to the minimum  

 

 

 

position.  This should be done prior to each time the base is turned on or off.  This will extend bulb  

 

 

 

life.

 

 

 

 

2.

 

Insert power plug into base of microscope.  Then insert plug into outlet.

 

 

3. Push rocker switch at back of the base to on position.

 

 

4. Rotate intensity dial on illuminator base until the image is illuminated.

 

 

5. Adjust intensity of light to match requirements of objective and specimen slide.

 

B. 

Interpupillary Adjustment of Viewing Head

 

 

1. Looking through microscope, adjust distance between the two eyepieces by grasping the eyetubes and  

 

 

moving them either closer or farther apart.

 

 

2. When a full field of view is observed throught both tubes, and images blend into one, interpupillary   

 

 

distance is corrected for your eyes.  Check the interpupillary scale and note index reading for future

 

 

reference, in case other users will be changing this adjustment from time to time.

 

 

3. Adjust the diopter scales, located on each eyetube, to the same numerical value as indicated on the   

 

 

interpupillary scale.  This must be done in order to maintain parfocality of objective lenses.  If interpupillary  

 

 

distance is changed, adjust eyepiece diopters accordingly.

 

C. 

Focusing the Microscope

 

 

1. Position the 4x objective lens into the optical path, making sure that lens clicks into its proper position.  

Содержание U-1B

Страница 1: ...Richter Optica 6122 Innovation Way Carlsbad California 92009 800 942 0528 US TOLL FREE 760 438 0528 INTERNATIONAL Instructions for Models U 1B U 1T U 1D U 1LCD Universty Student Biological Microscope...

Страница 2: ...ocus Diopter Adjustment Ring Condenser Control Knob X Axis Control Knob Y Axis Control Knob Set Screws to secure specimen holder to stage Head Locking Screw Interpupillary Adjustment Scale Iris diaphr...

Страница 3: ...ach time the base is turned on or off This will extend bulb life 2 Insert power plug into base of microscope Then insert plug into outlet 3 Push rocker switch at back of the base to on position 4 Rota...

Страница 4: ...image Smaller apertures will deliver contrast to the image However closing aperture too much will reduce resolution Experimentation is the best method of determining the correct opening of the diaphr...

Страница 5: ...ent is clean If immersion oil or specimen material of any kind is evident use a cotton swab dipped in a small amount of Windex to clean all foreign material from objective lens surface Such material w...

Страница 6: ...rs The voltage is too high Incorrect lamp used Lamp not properly inserted into socket Lamp about to burn out Fuse holder not locked into proper position Loose connection at AC outlet Adjust intensity...

Страница 7: ...ht too bright Objective lens dirty Eyepiece lens dirty Slide upside down Cover slip on specimen slide too thick Too much light Condenser lens dirty Rack stop not set at a proper position Move revolvin...

Отзывы: