background image

2

 

 

2. Place standard specimen slide (cover slip up) on top of stage surface.

 

 

 

a. Swing moveable finger on slide holder outward.  Place specimen slide against fixed side of slide  

 

 

 

   holder. Slowly release moveable finger until it makes contact with the specimen slide.

 

 

3. Rotate coarse focusing controls until the specimen comes into focus.

 

 

4. Adjust fine focus controls until specimen is in sharp focus.

 

 

5. Adjust diopter for difference in eyesight.

 

 

 

a. Using right eye, peer into the right eyepiece tube.  Adjust sharpness of image by turning diopter  

 

 

 

    adjustment located on left eyepiece tube.

 

 

6. Adjust the aperture (opening) of the iris diaphragm.

 

 

    Iris diaphragm should not be used to control the brightness of illumination, use light intensity control  

 

 

    knob to adjust light level.  Iris diaphragms are designed to help achieve high resolution of specimen and  

 

 

    provide contrast in the image. Smaller apertures will deliver contrast to the image.  However, closing   

 

 

    aperture too much will reduce resolution.  Experimentation is the best method of determining the    

 

 

    correct opening of the diaphragm.  Some suggested openings for the iris diaphragm are:

 

 

7. Changing magnification.

 

 

 

a. Rotate revolving nosepiece to postition 10x objective into optical path.

 

 

 

b. This microscope has been parfocalized, which allows changes from one objective to another  

 

 

 

    while requiring only a slight  adjustment of the fine focus controls.

 

 

 

c. When changing to the 40x and 100x objective lens, care must be exercised in order to prevent  

 

 

 

   damaging the front lens element and specimen slide.

 

 

 

d. In order to obtain the maximum resolution of the 100x oil immersion lens, it is necessary to    

 

 

 

   apply immersion oil between the cover glass of slide and the front lens of the objective.

 

 

 

 

1.) Use a very small amount of immersion oil.  Only the tip of the lens should ever come in  

 

 

 

 

      contact with the oil.

 

 

Objective

4x

10x

40x

100x

Diaphragm  Opening

1/8 open

1/8 to 1/4 open

1/4 to 1/2 open

1/2 to 3/4 open

Содержание U-1B

Страница 1: ...Richter Optica 6122 Innovation Way Carlsbad California 92009 800 942 0528 US TOLL FREE 760 438 0528 INTERNATIONAL Instructions for Models U 1B U 1T U 1D U 1LCD Universty Student Biological Microscope...

Страница 2: ...ocus Diopter Adjustment Ring Condenser Control Knob X Axis Control Knob Y Axis Control Knob Set Screws to secure specimen holder to stage Head Locking Screw Interpupillary Adjustment Scale Iris diaphr...

Страница 3: ...ach time the base is turned on or off This will extend bulb life 2 Insert power plug into base of microscope Then insert plug into outlet 3 Push rocker switch at back of the base to on position 4 Rota...

Страница 4: ...image Smaller apertures will deliver contrast to the image However closing aperture too much will reduce resolution Experimentation is the best method of determining the correct opening of the diaphr...

Страница 5: ...ent is clean If immersion oil or specimen material of any kind is evident use a cotton swab dipped in a small amount of Windex to clean all foreign material from objective lens surface Such material w...

Страница 6: ...rs The voltage is too high Incorrect lamp used Lamp not properly inserted into socket Lamp about to burn out Fuse holder not locked into proper position Loose connection at AC outlet Adjust intensity...

Страница 7: ...ht too bright Objective lens dirty Eyepiece lens dirty Slide upside down Cover slip on specimen slide too thick Too much light Condenser lens dirty Rack stop not set at a proper position Move revolvin...

Отзывы: