2.2 Warnings, Cautions and Adverse Reactions

 WARNINGS:

1.  This device should be used only under the continued supervision  

 

of a licensed physician or practitioner.

2.  The long-term effects of chronic electrical stimulation are  

 

 

unknown. Electrical stimulation devices 

DO NOT

 have any curative value.

3.  TENS is a symptomatic treatment and, as such, suppresses the    

 

sensation of pain, which would otherwise serve as a protective    

 mechanism.

4.  Safety has not been established for the use of therapeutic 

 

 

electrical stimulation during pregnancy. 

DO NOT

 use during  

 

 

pregnancy unless directed by your physician.

5.  Electrical stimulation is not effective for pain of central origin, such  

 

as a headache.

6.  Electrical monitoring equipment (such as ECG monitors and ECG  

 

alarms) may not operate properly when electrical stimulation is in use.

7.  Stimulation should not be applied over the carotid sinus nerves, 

 

particularly in patients with a known sensitivity to the carotid sinus reflex.

8.  Stimulation should not be applied over the neck or mouth. Severe  

 

spasm of the laryngeal and pharyngeal muscles may occur and the  

 

contraction may be strong enough to close the airway or cause    

 

difficulty in breathing.

9.  Stimulation should not be applied transthoracically. Introduction of  

 

electrical current into the heart may cause cardiac arrhythmias.

10.  Stimulation should not take place while the user is connected to   

 

high-frequency surgical equipment, it may cause burn injuries on 

 

the skin under the electrodes, as well as problems with the stimulator.

11. 

DO NOT

 use the stimulator in the vicinity of shortwave or microwave  

 

therapy equipment, since this may affect the output power of the  

 stimulator.

12. 

NEVER

 use in environments with high humidity such as in the    

 

bathroom or when having a bath or shower.

13.  Caution should be used in applying electrical stimulation to  

 

 

patients suspected of having heart disease. Further clinical data is  

 

needed to show there are no adverse results. 

14. 

NEVER

 use near the heart. Stimulation electrodes should 

NEVER

 be  

 

placed anywhere on the front of the thorax (marked by ribs and   

 

breastbone), take extreme caution not to place near or on the 

 

two large pectoral muscles. Here it can increase the risk of  

 

 

ventricular fibrillation and lead to cardiac arrest.

15.  Electrodes should not be placed over the eyes, in the mouth, near  

 

the genitals or internally.

Содержание Tens 3000

Страница 1: ...INSTRUCTION MANUAL Tens 3000 TENS 3000 TENS 3000...

Страница 2: ...ice Amendments may however be published in new editions of this manual All Right Reserved Rev V2 2019 United States Federal Law restricts this device to sale by or on the order of a physician or licen...

Страница 3: ...ms 5 INSTRUCTIONS FOR USE ________________________________ 14 5 1 Battery 5 2 Connect Electrodes to Lead Wires 5 3 Electrode Options 5 4 Adjusting the Controls 5 5 Turn ON 6 CLEANING AND CARE ________...

Страница 4: ...cessary warning signal of trauma or malfunction in the body nature may have gone too far in its design Aside from its value in diagnosis long lasting persistent pain serves no useful purpose Pain does...

Страница 5: ...individual patient mode selected for therapy and the type of pain In many patients the reduction or elimination of pain lasts longer than the actual period of stimulation sometimes as much as three t...

Страница 6: ...een diagnosed 2 This device should not be used when cancerous lesions are present in the treatment area 3 Stimulation should not be applied over swollen infected inflamed areas or skin eruptions e g p...

Страница 7: ...es may occur and the contraction may be strong enough to close the airway or cause difficulty in breathing 9 Stimulation should not be applied transthoracically Introduction of electrical current into...

Страница 8: ...iagnosed pain syndromes until consulting a physician 8 Patients with an implanted electronic device such as a cardiac pacemaker implanted defibrillator or any other metallic or electronic device shoul...

Страница 9: ...ipment which sends electrical pulses to your body 23 DO NOT confuse the electrode cables and contacts with your headphones or other devices and DO NOT connect the electrodes to other devices 24 DO NOT...

Страница 10: ...n irritation occurs discontinue use and consult your physician Note Always use electrodes that are legally marketed and sold in the United States under 510K guidelines 2 If the stimulation levels are...

Страница 11: ...QTY 1 TENS Device Item DT3002 1 each 2 Pair of Lead wires Item WW3005 2 pk 3 2 x 2 Self Adhesive electrodes Item EP2020WC2 INTM 4 pk 4 9V Battery Type 6F22 Item TA9050 I 1 each 5 Instruction manual 1...

Страница 12: ...TENS Mode Waveform Asymmetrical Biphasic Square Pulse Pulse Amplitude Adjustable 0 80mA at 500 ohm Load each channel 1mA Step Pulse Width Adjustable from 30 260 s Pulse Rate Adjustable from 2 to 150 H...

Страница 13: ...13 4 3 The Waveforms of the Stimulation Programs Burst Normal Modulation 10 pulses per burst...

Страница 14: ...ribbon to pop out easily and then insert a new 9V battery according to the above steps 2 to 5 5 1 2 Disposal of Battery Dispose of used batteries according to the current federal state and local regu...

Страница 15: ...proceeding to this step Connection of the electrodes is made with two lead connector Both intensity controls must be at the Off position Electrodes must be pressed firmly on the skin 5 3 Electrode Opt...

Страница 16: ...odes from the skin while the device is turned on You will feel an uncomfortable electrical shock 4 It is recommended that at a minimum 1 5 x 1 5 self adhering square electrodes are used at the treatme...

Страница 17: ...green whenever the electronics of the device create a current impulse Due to the capacity of the human eye the illumination of the lamp can only be recognized up to a frequency of approximately 30 Hz...

Страница 18: ...position white markings on the housing match up with the OFF dot on each control the device is switched off By turning clockwise to hear feel a click and then stop this means the appropriate channel i...

Страница 19: ...during application To prevent this apply electrodes from center outward avoid stretching over the skin 5 To minimize pulling stress tape extra lengths of lead wires to the skin in a loop to prevent t...

Страница 20: ...e and insert them into the resealable bag and store in a cool dry place 6 4 Cleaning the Lead Wires Clean the lead wires by wiping them with a damp cloth Coating them lightly with talcum powder will r...

Страница 21: ...d wires old worn damaged Replace Stimulation is uncomfortable Intensity is too high Decrease intensity Electrodes are too close together Reposition the electrodes Damaged or word electrodes or lead wi...

Страница 22: ...n becomes red and or you feel a stabbing pain Using the electrodes on the same site every time Reposition the electrodes If at any time you feel pain or discomfort stop use immediately The electrodes...

Страница 23: ...attery to avoid battery leaking 2 Store the device in a cool well ventilated place 3 NEVER place any heavy objects on the device Used fully discharged batteries must be disposed of in a specially labe...

Страница 24: ...ons CISPR 11 Group 1 The device uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emi...

Страница 25: ...ctrical fast transient burst IEC 61000 4 4 2 kV for power supply Not applicable Main power quality should be that of a commercial or hospital environment 1 kV for input output lines Surge IEC 61000 4...

Страница 26: ...I At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption an...

Страница 27: ...distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitter...

Страница 28: ...al and should not be disposed of with household waste after their useful life Help us to protect the environment and save resources and take this device to the ap propriate collection points Please co...

Страница 29: ...anufacturer to the consumer or during transport to the retailer Accessories which are subject to normal wear and tear Liability for direct or indirect consequential losses caused by the unit is exclud...

Страница 30: ...and return freight charges will apply All returns of dropshipped items are subject to a restocking fee as well as inbound and outbound freight charges 7 Returns will not be accepted on items that are...

Страница 31: ...Manufactured for 3287 19 06 A 2019 Compass Health Brands Corp...

Отзывы: