Communications
7.5.1 Input Command Composition 2
This is composed of 15 characters including the terminator (CR=0xDH/LF=0xAH).
C3 has a 10 digit maximum (including the po/-, comma and point). Do not include the measuring unit.
Example: Upper limit input 1200.00g: LA,1200.00
Preset tare input 1000.00g: PT,1000.00
Interval time input 12:34:56: IA,12,34,56 (marked off by commas)
The Input command can be entered when Weighing mode, Percentage mode, Counting mode or Multiplied by Coefficient mode
are being used. If input when in an other modes, the outputs an abnormal response.
If the input value is not correct, the balance outputs an abnormal response.
When the normal response, the preset tare value is input in 321 PRESET 1 and the balance operates Preset tare.
If the input value is 0 at Preset tare setting value command, the preset tare operation is canceled
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
C1
C2
,
C3
C3
C3
C3
C3
C3
C3
C3
C3
C3
CR
LF
C1
C2
Code (C1)
Code (C2)
Description
C3
Response
A00/Exx
format
ACK/NAK
format
L
A
0x4C
0x41
Upper limit value setting
Numeric value setting
A00:
Normal
response
E01:
Abnormal
response
ACK:
Normal
response
NAK:
Abnormal
response
L
B
0x4C
0x42
Lower limit value setting
Numeric value setting
L
C
0x4C
0x43
Reference value setting
Numeric value setting
Table 7-8. Comparator Setting Command
C1
C2 Code (C1) Code (C2)
Description
C3
Response
A00/Exx format A00/Exx format
P
T
0x50
0x54
Preset tare value setting
Numeric value setting A00:
Normal response
E01:
Abnormal
response
ACK:
Normal response
NAK:
Abnormal
response
Table 7-9. Preset Tare Value Setting Command
C1
C2
Code
(C1)
Code
(C2)
Description
C3
Response
A00/Exx
format
A00/Exx
format
I
A
0x49
0x41
Interval (output) time setting Numeric value setting A00:
Normal
response
E01:
Abnormal
response
ACK:
Normal
response
NAK:
Abnormal
response
Table 7-10. Interval (Output) Time Setting Command