Operation
3.6 Zero-Point Adjustment
A zero-point adjustment is an adjustment of the scale’s zero value with all weight removed from the scale. The scale must be
stable and within the zero range to perform this.
On Legal for Trade balances, this setting is read only, the wait time cannot be changed.
Use the following steps to do a zero-point adjustment.
1. Ensure the weigh pan is empty.
2. Press
,
ZERO TARE
displays.
3. 0000000 displays after zero is complete.
The stability wait time can be set in the Applications menu in
.
3.6.1 Zero-Point Adjustment Limits
The
Zero Point
limit is the threshold at which the operator can press
and have it zero.
There is an upper and lower threshold, outside of that, the balance cannot be zeroed.
Weight above 1.5% of capacity will be processed as a tare but a recalibration would be needed to adjust a zero point
outside of this.
3.7 Tare Value
To weigh using a container, the container weight should be subtracted from the total value. Then results displayed are the
weight of the product only.
A Tare Value is included in the total maximum capacity.
If a tare exceeds the zero point adjustment range (
) on initial power up, the tare subtraction is executed.
1. Place the empty container on the balance. The weight of the container displays.
2. Press
,
ZERO TARE
displays.
3. 0000000 displays after zero is complete.
4. Place the product to be weighed in the container. The net weight displays.
5. Remove the product and container from the balance.
6. Press
to remove the tare.
Weight above 1.5% of capacity will be processed as a tare but a recalibration would be needed to adjust a zero point
outside of this.
3.7.1 Preset Tare
When a tare weight is already known, the tare subtraction can be configured as a
Preset Tare
parameter. Five preset tare
values can be stored.
To use a preset tare:
1. Press
.
displays.
2. Press
until
3 USER INFO
displays.
3. Press
and
31 PT MODE
and the current setting displays.
Model
Lower Limit (g) Upper Limit (g)
TA Plus - 124
-1.8
1.8
TA Plus - 224
-3.3
3.3
Table 3-3. Zero-Point Adjustment Limits