5.4.2 Input Preset Tare Value
There are two ways of inputting a preset tare weight value
• Actual value setting method (onW): Weighing a sample with a scale and then making it a setting value
• Numeric value setting method (NUM): Inputting a setting value directly via key operation
To enter a preset tare:
1. Press
.
displays.
2. Press
or
to scroll to
USER INFO
.
3. Press
.
31 PT MODE
displays.
4. Press
or
to scroll to
31 PT INPUT
.
5. Press
.
PRESET 1
displays.
6. Press
or
to select desired preset tare (1-5).
7. Press
.
SET PRESET onW NUM
displays.
8. Select
for
onW
or
for
NUM
.
• If
onW
is selected, place the container on the scale, when weight is stable, press
to save
• If
NUM
is selected use
or
to enter known value, press
to save
9. Press
to return to operation display.
NET PT
display in the upper right.
5.4.3 Set the Discrimination Value of Comparator Function
There are two ways of inputting a reference value and upper and lower limit values
• Actual value setting method (onW): Weighing a sample with a scale and then making it a setting value
• Numeric value setting method (NUM): Inputting a setting value directly via key operation
The discrimination is performed according to the following criteria:
Absolute Value
The discrimination is performed based on the upper and lower limit values that have been set in advance.
Relative Value
A reference numeric value is set in advance, and the discrimination is performed based on the range defined by the upper and
lower limit values that have been set for the reference numeric value.
Example:
Two-point (upper and lower limits) setting, Reference value = 100.0000 g
Lower limit value = 90.0000 g, Upper limit value = 120.0000 g
Discrimination
Method
Reference value Lower limit value
Upper limit value
100.0000 g
90.0000 g
120.0000 g
Absolute value
90.0000 g
120.0000 g
Relative value
100.0000 g
-10.0000 g
20.0000 g
Table 5-7. Relative Value Example