RS-232 Communications

17

7.0

RS-232 Communications

The scale comes with an RS-232 port which enables weight data to be transmitted to other equipment, such as a 

computer or printer. The RS-232 cable with DB-9 connector (PN 100719) is available from Rice Lake Weighing 

System. Figure 4-1 on page 7 shows where the RS-232 connection is.
The RS-232 parameters are 9600 baud (selectable in the programming mode), 8 data bits, 1 stop bit, no parity 

and no handshaking.
There are three methods of communication:

Pushbutton keypad print

Standard remote protocol

Escape protocol

7.1

Pushbutton Keypad Print

With a stable, in-range weight, press and hold the 

Kg-Lb/Print

 key for at least three seconds, or until the scale 

emits two quick beeps. Note that if the scale does not beep after five seconds, then release the button as the 

weight was either in motion, or out of range.

If displaying weight and not BMI, the scale will send out the following 21 character string:

xxxxxxxxx<SP>uu<SP>mmmmm<SP><CR><LF>
Where:

xxxxxxxxx is the weight with decimal point and " - " sign, if negative uu is the unit (lb or kg). 
mmmmm is the mode (gross or net)

Examples:

-10 Lb net = <SP><SP><SP><SP>-10.0<SP>lb<SP><SP>Net<SP><SP><SP><CR><LF>
10 Lb gross = <SP><SP><SP><SP><SP>-10.0<SP>lb<SP>Gross<SP><CR><LF>

In BMI mode (displaying the BMI value), the scale will send out the following data:

GROSS WEIGHT       215.0 LB
TARE WEIGHT               0.0 LB
NET WEIGHT             215.0 LB
PATIENT HEIGHT      6-01.0 FT
PATIENT BMI                28.4

7.2

Standard Remote Protocol (configuration option #9 set to 1)

When connected to a computer, there are five commands that can be used in the standard remote protocol to 

communicate with the scale. They are:

t - tare the scale. If in gross mode, will tare and go to net mode. If in net mode, will remove tare and 

return to gross mode.

w - the scale sends the actual weight to the computer.

i - the scale sends the software ID of the scale.

z - the scale will be set to zero (0.0) if possible.

p - the scale sends the same data as a pushbutton keypad print. 

The format of the returned data will be the same as noted under the pushbutton keypad print. Note that the w and 

p commands will not return a value if the scale is in motion, or displaying an invalid weight. 

7.3

ESC Protocol (configuration option #9 set to 0)

The ESC Protocol differs from the standard protocol and enables weight and unit of measure data to be 

transmitted for full integration into electronic medical records or for diagnostic testing of the battery life, load 

cells, etc. The scale will only transmit data upon receiving the proper command set. 
You can test the command set and review the scales’ response using either PROCOMM Plus or the 

Dietary/Fitness files found on our web site, 

www.ricelake.com/health

. If using PROCOMM Plus, we 

recommend you set up "hot" keys for the commands. Refer to Table 7-1 for a listing of those commands.

Содержание Summit SB-1000

Страница 1: ...106992 Summit SB 1000 Bariatric Wheelchair Scale Installation Operation Manual...

Страница 2: ......

Страница 3: ...0 2 Using the Body Mass Index BMI Function 14 6 0 3 Using the Tare Function 15 7 0 RS 232 Communications 16 7 1 Sample and Explanation of ESC Protocol 19 8 0 Service Information 21 8 1 Troubleshootin...

Страница 4: ...ii Summit SB 1000 Bariatric Wheelchair Scale Installation Manual...

Страница 5: ...ear each corner eliminating the possibility of cable damage in portable applications One threaded hole located in the center of the deck is another useful feature for portable applications This allows...

Страница 6: ...e water may damage a scale not meant for a washdown environment The interface cable between the scale and the indicator must be protected against crushing cutting or moisture damage If the chosen site...

Страница 7: ...ment be required in the future cable installation instructions are shown below The junction box is easily accessible through an access plate located on the side of the Summit SB 1000 Use the following...

Страница 8: ...ps Access ramps for Summit SB 1000 Bariatric Wheelchair Scales are designed to bolt to the floor with built in mounting plates that attach to the scale feet When used with access ramps side movement o...

Страница 9: ...vel of the deck with a spirit level The deck must be level within 1 4 3 2 Corner Correction Trimming The assembled Summit SB 1000 scale is delivered with the junction box corner trimmed Corner trimmin...

Страница 10: ...fect the reference cell output especially in smaller scale decks Rezero the indicator and repeat the test until all corners read within 1 of the test weight used 3 3 Calibration Procedure Refer to the...

Страница 11: ...ion Site with Back Plastic Cover in Place The Summit SB 1000 Scale is capable of running its internal sealed lead acid rechargeable battery if no additional power source is available Battery life is a...

Страница 12: ...mode the configuration enable jumper must be removed Access to that jumper is gained by removing the back cover of the indicator Remove the four back cover retaining screws as shown on the left side...

Страница 13: ...us parameters can be set up while in configuration mode Table 4 1 Programming Mode Menu Structure Numeric Data Entry Use the Kg Lb key to change the numeric data entry while setting up the various con...

Страница 14: ...aximum allowed weight Any weight above this value will cause StOP to appear on the display while in the weighing mode FULL XX XXX Default 600 Lb Note This value is dependent on the model you have Refe...

Страница 15: ...measure UOM in calibration and programming 0 Kg 1 Lb Option 2 allows the scale to work only in Kg OP2 0 Hold the Kg Lb key for browsing through the options Double click on the Kg Lb key to change the...

Страница 16: ...options value or to advance to the next step press the REWEIGH key 0 Disable 1 Enable Note Leave set to 0 Option 8 allows you to choose between 9 volts and 6 volts OP8 1 Hold the Kg Lb key for browsin...

Страница 17: ...pean 1 default 3 European 2 default 9 Used if OP1 through OP10 have been changed to something other than the factory default settings Note Always set to 1 or 9 for US indicators To advance to the next...

Страница 18: ...To advance to the next step press the REWEIGH key 3 Sets the value of the calibration weight you are going to use for calibrating the span value of the scale LOAd XXX X Use the numeric data entry expl...

Страница 19: ...ting BMI is one of the best methods for population assessment of overweight and obesity Because calculation requires only height and weight it is inexpensive and easy to use for clinicians and for the...

Страница 20: ...continue to weigh patients who are carrying the same tare weight For example when using the same wheelchair for weighing more than one patient 6 To cancel the tare weight press and hold the TARE key...

Страница 21: ...following data GROSS WEIGHT 215 0 LB TARE WEIGHT 0 0 LB NET WEIGHT 215 0 LB PATIENT HEIGHT 6 01 0 FT PATIENT BMI 28 4 7 2 Standard Remote Protocol configuration option 9 set to 1 When connected to a...

Страница 22: ...he scale is overloaded the scale will return the value of 999 9 Units N Nc This indicates in which unit the values have been taken m metric c constitutional End of Packet EOP E E This indicates that t...

Страница 23: ...Inbar ini 5 Click on STRINGS and the following screen appears Figure 7 2 Strings to be Used in the RS 232 Transmission This screen is showing that the function keys are already pre programmed with com...

Страница 24: ...3 Figure 7 3 HEX Screen Example Examples of what you would see in the ASCII screen are shown in Figure 7 4 Figure 7 4 ASCII Screen Example 7 4 Sample and Explanation of ESC Protocol When the scale mea...

Страница 25: ...t command and will look like this If the battery is okay the scale will reply with the following value If the battery is critically low it will reply with 01 6OJU PG FBTVSF FUSJD 4 3 4 8 4 N 4 3FBEJOH...

Страница 26: ...Place the patient back on the scale Press the REWEIGH button again Consistently high or low weights Indicator not properly adjusted to zero Zero the indicator according to the indicator manual Platfo...

Страница 27: ...the scale exceeds the capacity of the scale Remove the excess weight and use the scale according to manufacture s specs The display shows LO Bat message The battery is low Recharge the battery The dis...

Страница 28: ...e unit cools to room temperature Step Function Display Available Parameters 1 Turn the scale on While StArt is displayed press and hold the REWEIGH key This enters into the test mode of the scale StAr...

Страница 29: ...box according to the wiring diagram in Figure 8 3 Note that in Figure 8 3 both the scale and the junction box are viewed from the bottom To verify correct load cell junction box terminal matching see...

Страница 30: ...er correction trimming and calibration is necessary after load cell replacement Follow instruction in Sections 3 2 on page 5 8 3 1 Load Cell Wiring to Junction Box The four load cells are each wired t...

Страница 31: ...ments The supporting surface for the four feet of the scale must be level within 1 4 inch of horizontal Safe Static Overloading Capacity Maximum 150 of scale capacity Nominal Scale Height 3 0 inch 76...

Страница 32: ...confirms that the nonconformity actually exists and was not caused by accident misuse neglect alteration improper installation improper repair or improper testing RLWS shall be the sole judge of all...

Отзывы: