Indicator Installation and Configuration

13

4.4

Reset To Factory Defaults

The configuration parameters may be reset to factory defaults while in the configuration mode. To do so, enter 

the configuration mode. Press the 

REWEIGH

 key once to advance to displaying the software version (ID <-> 

11305). Press and hold the 

BMI

 key until 

=DEF=

 is displayed. At this point, if you wish to perform the default 

function, press the 

REWEIGH

 key and the scale will show 

DONE

, then it will restart. If you do not want to perform 

the default function, press the 

ON/OFF

 key.

After performing the default function, the scale will require re-calibration. In addition, the scale will be set 

up to a capacity of 600 lbs. The capacity varies by model - refer to the serial label on your scale and 

reconfigure the capacity (FULL) and display divisions (rOUnd) as necessary in the configuration menu.

 

15

Set Option defaults

Set < - > 

1

To change, press the 

Kg-Lb

 key.

0 - default
Options = (OPx)

1 = US default

 (US defaults are 

OP1-1, OP2-0, OP3-0, OP4-0, OP5-0, 
OP6-1, OP-7-0, OP8-1, OP9-1, 
OP10-1).
2 = European 1 default
3 = European 2 default
9 = Used if OP1 through OP10 have 
been changed to something other than 
the factory default settings.

Note:

 Always set to 1 or 9 for US 

indicators. To advance to the next 
step, press the 

REWEIGH

 key.

16

Determines the 
automatic shut off 
time when the scale is 
not in use. Options are 
between one and 20 
minutes.
This is used when the 
unit is battery 
operated.

t-OFF  <-> 

5

Press and hold the 

Kg-Lb

 key to scroll 

through the furthest right hand digit on 
the display (0-9). If you want a value 
from 10 to 19, with the display sitting 
at 1, double press the 

Kg-Lb

 key and 

10 appears. Press and hold the 

Kg-Lb

 

key to begin scrolling 10-19. If you 
want a value of 20, with the display 
sitting at 2, double press the 

Kg-Lb

 

key and 20 will appear.
To advance to the next step, press the 

REWEIGH

 key.

17

StArt or SAvE

If no changes were made in the 
parameters the display will 
automatically show 

StArt 

and then 

returns to normal weighing mode.
If a parameter was changed, the 
display will show 

SAvE.

 To save 

changes made, press the 

REWEIGH

 

key. To return to weigh mode without 
saving changes, press the 

Kg-Lb

 key.

18

Con En

Replace the configuration jumper.

Step

Function

Display

Available Parameters

Table 4-2. Configuration Mode Menu

Содержание Summit SB-1000

Страница 1: ...106992 Summit SB 1000 Bariatric Wheelchair Scale Installation Operation Manual...

Страница 2: ......

Страница 3: ...0 2 Using the Body Mass Index BMI Function 14 6 0 3 Using the Tare Function 15 7 0 RS 232 Communications 16 7 1 Sample and Explanation of ESC Protocol 19 8 0 Service Information 21 8 1 Troubleshootin...

Страница 4: ...ii Summit SB 1000 Bariatric Wheelchair Scale Installation Manual...

Страница 5: ...ear each corner eliminating the possibility of cable damage in portable applications One threaded hole located in the center of the deck is another useful feature for portable applications This allows...

Страница 6: ...e water may damage a scale not meant for a washdown environment The interface cable between the scale and the indicator must be protected against crushing cutting or moisture damage If the chosen site...

Страница 7: ...ment be required in the future cable installation instructions are shown below The junction box is easily accessible through an access plate located on the side of the Summit SB 1000 Use the following...

Страница 8: ...ps Access ramps for Summit SB 1000 Bariatric Wheelchair Scales are designed to bolt to the floor with built in mounting plates that attach to the scale feet When used with access ramps side movement o...

Страница 9: ...vel of the deck with a spirit level The deck must be level within 1 4 3 2 Corner Correction Trimming The assembled Summit SB 1000 scale is delivered with the junction box corner trimmed Corner trimmin...

Страница 10: ...fect the reference cell output especially in smaller scale decks Rezero the indicator and repeat the test until all corners read within 1 of the test weight used 3 3 Calibration Procedure Refer to the...

Страница 11: ...ion Site with Back Plastic Cover in Place The Summit SB 1000 Scale is capable of running its internal sealed lead acid rechargeable battery if no additional power source is available Battery life is a...

Страница 12: ...mode the configuration enable jumper must be removed Access to that jumper is gained by removing the back cover of the indicator Remove the four back cover retaining screws as shown on the left side...

Страница 13: ...us parameters can be set up while in configuration mode Table 4 1 Programming Mode Menu Structure Numeric Data Entry Use the Kg Lb key to change the numeric data entry while setting up the various con...

Страница 14: ...aximum allowed weight Any weight above this value will cause StOP to appear on the display while in the weighing mode FULL XX XXX Default 600 Lb Note This value is dependent on the model you have Refe...

Страница 15: ...measure UOM in calibration and programming 0 Kg 1 Lb Option 2 allows the scale to work only in Kg OP2 0 Hold the Kg Lb key for browsing through the options Double click on the Kg Lb key to change the...

Страница 16: ...options value or to advance to the next step press the REWEIGH key 0 Disable 1 Enable Note Leave set to 0 Option 8 allows you to choose between 9 volts and 6 volts OP8 1 Hold the Kg Lb key for browsin...

Страница 17: ...pean 1 default 3 European 2 default 9 Used if OP1 through OP10 have been changed to something other than the factory default settings Note Always set to 1 or 9 for US indicators To advance to the next...

Страница 18: ...To advance to the next step press the REWEIGH key 3 Sets the value of the calibration weight you are going to use for calibrating the span value of the scale LOAd XXX X Use the numeric data entry expl...

Страница 19: ...ting BMI is one of the best methods for population assessment of overweight and obesity Because calculation requires only height and weight it is inexpensive and easy to use for clinicians and for the...

Страница 20: ...continue to weigh patients who are carrying the same tare weight For example when using the same wheelchair for weighing more than one patient 6 To cancel the tare weight press and hold the TARE key...

Страница 21: ...following data GROSS WEIGHT 215 0 LB TARE WEIGHT 0 0 LB NET WEIGHT 215 0 LB PATIENT HEIGHT 6 01 0 FT PATIENT BMI 28 4 7 2 Standard Remote Protocol configuration option 9 set to 1 When connected to a...

Страница 22: ...he scale is overloaded the scale will return the value of 999 9 Units N Nc This indicates in which unit the values have been taken m metric c constitutional End of Packet EOP E E This indicates that t...

Страница 23: ...Inbar ini 5 Click on STRINGS and the following screen appears Figure 7 2 Strings to be Used in the RS 232 Transmission This screen is showing that the function keys are already pre programmed with com...

Страница 24: ...3 Figure 7 3 HEX Screen Example Examples of what you would see in the ASCII screen are shown in Figure 7 4 Figure 7 4 ASCII Screen Example 7 4 Sample and Explanation of ESC Protocol When the scale mea...

Страница 25: ...t command and will look like this If the battery is okay the scale will reply with the following value If the battery is critically low it will reply with 01 6OJU PG FBTVSF FUSJD 4 3 4 8 4 N 4 3FBEJOH...

Страница 26: ...Place the patient back on the scale Press the REWEIGH button again Consistently high or low weights Indicator not properly adjusted to zero Zero the indicator according to the indicator manual Platfo...

Страница 27: ...the scale exceeds the capacity of the scale Remove the excess weight and use the scale according to manufacture s specs The display shows LO Bat message The battery is low Recharge the battery The dis...

Страница 28: ...e unit cools to room temperature Step Function Display Available Parameters 1 Turn the scale on While StArt is displayed press and hold the REWEIGH key This enters into the test mode of the scale StAr...

Страница 29: ...box according to the wiring diagram in Figure 8 3 Note that in Figure 8 3 both the scale and the junction box are viewed from the bottom To verify correct load cell junction box terminal matching see...

Страница 30: ...er correction trimming and calibration is necessary after load cell replacement Follow instruction in Sections 3 2 on page 5 8 3 1 Load Cell Wiring to Junction Box The four load cells are each wired t...

Страница 31: ...ments The supporting surface for the four feet of the scale must be level within 1 4 inch of horizontal Safe Static Overloading Capacity Maximum 150 of scale capacity Nominal Scale Height 3 0 inch 76...

Страница 32: ...confirms that the nonconformity actually exists and was not caused by accident misuse neglect alteration improper installation improper repair or improper testing RLWS shall be the sole judge of all...

Отзывы: