14
Summit SB-1000 Bariatric Wheelchair Scale Installation Manual
5.0
Scale Calibration
Before you can calibrate the scale, verify and set all scale parameters which are noted in “Indicator Installation
and Configuration” on page 7.
Remove the configuration/calibration jumper (see page 8). Turn the scale on. While
StArt
is displayed, press and
hold both the
Kg-Lb
key and the
REWEIGH
keys until
CAL
is displayed.
To do a calibration on the scale, the scale must be turned off.
NOTE:
The calibration weight must be no less than 60 lb (28 kg) and no more than 300 lb (135 kg).
The following chart illustrates the calibration procedure
Step
Function
Display
Available Parameters
1
Turn the scale on.
While
START
is
displayed, press and
hold the
Kg-Lb
and
REWEIGH
keys until
CAL
is displayed.
Enters into calibration
mode
StArt
The scale automatically advances to
Step 2.
2
Cal mode entered
CAL
To advance to the next step, press the
REWEIGH
key.
3
Sets the value of the
calibration weight
you are going to use
for calibrating the
span value of the
scale.
LOAd <-> XXX.X
Use the numeric data entry (explained in
“Numeric Data Entry” on page 9) to set a
calibration weight.
To advance to the next step, press the
REWEIGH
key.
4
CLEAr
Clear the platform and be sure of the
scale’s stability.
To advance to the next step, press the
REWEIGH
key.
5
Calibrate Zero
- - - - -
PUT<->XXXX
Place the requested weight on the scale.
This will display for a few seconds.
To advance to the next step, press the
REWEIGH
key.
6
Calibrate Span
CAL
FAC tOr <-> X.XXX
This will be displayed for a few seconds
and shows the current calibration factor.
To advance to the next step, press the
REWEIGH
key.
7
SAvE
To save the new calibration, press the
REWEIGH
key.
To exit without saving the calibration
press the
ZERO
key.
8
dONE
The scale displays that it has saved that
calibration value and automatically
advances to the last step.
9
StArt
The scale reboots.
10
Con En
Replace the configuration jumper.
Table 5-1. Calibration Menu