EL232 XPCD Operation

19

4.0

EL232 XPCD Operation

Prior to powering up the EL232 XPCD, double check wiring and setup making sure all wiring is properly
terminated. 

Once power is applied, the EL232 XPCD enters the power on clear mode. All status switches are read at this time
and the unit is configured for baud rate, indicator type, and input signal type. The EL232 XPCD display goes
through a power up sequence which can provide valuable information about the operating condition and setup of
the unit. 

4.1

Lamp Test

The first display during power on clear is a lamp test.
This test lights all displays, decimal points, and
annunciators. Proper indication is “

8.8.8.8.8.8.

” with

five illuminated LEDs above the main display.

Figure 4-1. EL232 XPCD LED Display

4.2

Software Revision

The software revision number which is in the EL232
XPCD is displayed. The display shows 

rEV 2.6

 for

normal revision 2.6 program. This helps identify
specially modified software.

4.3

Current Baud Rate Setting

The EL232 XPCD next shows the current baud rate
setting it had read from switch SW1. The display
shows, 

bAUd12

 for 1200 baud, 

bAUd24

 for 2400

baud, 

bAUd48

 for 4800 baud, or 

bAUd96

 for 9600

baud. 

4.4

Input Signal Type

The display shows 

rS232

 for RS232 data setting or 

C LOOP

 for current loop input. Refer to Section 3.1

for further information on setting the correct input
data type.

4.5

Indicator Type

Finally, the EL232 XPCD shows the indicator type
which the EL232 XPCD is receiving. This is
displayed as 

Ind 1

 where the 1 can be an indicator

number from 1 to 12.  Refer to Table 4-1 for a
numerical list of indicator types.

From the power on clear sequence, the current status
of the display and the current setup information can be
verified. This can be done at any time by pressing the
reset switch (see Figure 3-9 on page 17) on the CPU
or by removing AC power from the unit. 

Numeric Display

Indicator Type

1

Condec, UMC

2

General Freedom

3

Electroscale 560

4

A N D Instruments

5

Weightronix 110, 120

6

Toledo Hi Speed data format

7

Flex-weigh DWM-4

8

Analogic

9

Masstron

10

Fairbanks

11

Streeter Richardson 9000

12

Cardinal 738

16

Generic Format

Table 4-1. Numeric Indicator Type

Содержание EL232 XPCD

Страница 1: ...16942 EL232 XPCD Explosion Proof Remote Display Installation Manual To be the best by every measure...

Страница 2: ......

Страница 3: ...t 14 3 2 Serial Data Input 14 3 2 1 RS232 C 14 3 2 2 20 mA Active Current Loop Input 15 3 2 3 20 mA Passive Current Loop Input 16 3 3 Configuration of the EL232 XPCD 17 3 3 1 Loss of Signal Display 18...

Страница 4: ...2 ES232 XPCD Installation Manual...

Страница 5: ...in this manual Improper specification installation or service of this equipment could result in personal injury or property damage Read all of the manual prior to installation Do not assume that all h...

Страница 6: ...DOUS AREA High temperature gases Cooling from conduction of heat along the threads There must be a minimum of five full threads of engagement of a 2 A 2 B fit for a Group C enclosure Group B increases...

Страница 7: ...that enclosure s rating Explosion HAZARDOUS AREA High temperature gases Cooling from expansion of gas Boyle s Gas Law The distance from the boundary of the inside of the enclosure to the gasket is kn...

Страница 8: ...4 EL232 XPCD Installation Manual...

Страница 9: ...323 4316 Weightronix WI 110 and WI 120 Toledo indicators with High speed data format Flex weigh DWM 4 Analogic AN5315 AN5316 AN5322 Masstron M 2500 and M 5000 Fairbanks 90 164 series Streeter Richards...

Страница 10: ...display digits and five LED annunciators Figure 1 1 EL232 XPCD Front Panel CLASS I DIVISION 2 GROUP C D CLASS II DIVISION 1 GROUP E F G TO PREVENT IGNITION OF HAZ ATMOS DISCONNECT O RE O PENING KEEP...

Страница 11: ...the enclosure It is recommended to shelter the outdoor installation by a shelter roof enclosure or covering Moist environments can also be controlled by the use of desiccants Contact your local dealer...

Страница 12: ...ain relief fitting to avoid pulling the wires out of the sealoff fittings in an accident Doing so would open a flame path to the enclosure and defeat its purpose Note The inside of an explosion proof...

Страница 13: ...he explosion proof enclosure is allowed to vent internal pressure appropriately the outer insulation must be removed back to the point ahead of entry into the safe area Multi conductor cables will pas...

Страница 14: ...amp with a blunt stick to expel any air bubbles Install and tighten filling opening cover Fittings requiring more than 16 oz of cement must be filled from a single mixture of cement and water Do not p...

Страница 15: ...ems Connect the black wire AC Hot to terminal strip J1 3 on the CPU board located in the bottom of the enclosure shown below Figure 2 3 J1 Location on CPU Board Connect the neutral wire white to termi...

Страница 16: ...lectrical way that data is transmitted The EL232 XPCD can read the data transmission strings sent by 12 of the more popular scale indicator families as shown in Table 3 1 It cannot read data from indi...

Страница 17: ...oop or RS232 using S3 10000100 1200 to 9600 See Figure 3 4 Generic smart receiver N A RS232 or current loop 1200 to 9600 Can receive serial data in a wide variety of intuitive formats Use when all els...

Страница 18: ...e connects the ground planes of the sending and receiving instruments RS232 should never be run more than 50 feet It may work but it may not work well or all the time Long RS232 data cables can act as...

Страница 19: ...current loop means that the instrument which sends the data also supplies the voltage to power the loop This voltage can be as small as 5 VDC to as much as 24 VDC When the loop goes through an optoco...

Страница 20: ...n external power supply or receiver This voltage can be a small as 5 VDC to as much as 24 VDC When the loop goes through an optocoupler a drop occurs This drop is about 2 VDC A current loop can power...

Страница 21: ...able indicators The power must be off when these values are changed If the power is on press the Reset switch on the CPU board to enter the new values see Figure 3 9 Figure 3 9 Reset Switch Location J...

Страница 22: ...transmitting to cannot properly receive the data change the 4N33 to a 4N25 opto isolators and the problem will normally disappear Figure 3 10 illustrates how to connect the current loop output of the...

Страница 23: ...ng The EL232 XPCD next shows the current baud rate setting it had read from switch SW1 The display shows bAUd12 for 1200 baud bAUd24 for 2400 baud bAUd48 for 4800 baud or bAUd96 for 9600 baud 4 4 Inpu...

Страница 24: ...SW1 07 is open the last value received is displayed continuously If this switch is closed the display again shows until readable data is again received This can be used to alert the operator that dat...

Страница 25: ...indicator incompatability mis wiring and mis programming both in the indicator and in the EL232 XPCD setup Double check the output of the meter and make sure something is being transmitted Use an osc...

Страница 26: ...ately 25 lb 11 3 kg Baud Rate 1200 2400 4800 and 9600 Interfaces 20 mA and RS232 standard The EL232 XPCD can communicate as shipped with twelve of the most common scale indicator families currently av...

Страница 27: ...accident misuse neglect alteration improper installation improper repair or improper testing RLWS shall be the sole judge of all alleged non conformities Such equipment has not been modified altered...

Страница 28: ......

Страница 29: ......

Страница 30: ...230 W Coleman St Rice Lake WI 54868 USA U S 800 472 6703 Canada Mexico 800 321 6703 International 715 234 9171 www ricelake com mobile m ricelake com 2011 Rice Lake Weighing Systems PN 16942 06 11...

Отзывы: