Calibration
© Rice Lake Weighing Systems
●
All Rights Reserved
19
5.3 Initial Calibration
Use this procedure only if the capacity and count-by (
d
) needs to be modified. The initial steps of the initial calibration will totally
erase user setup as well as any previous calibration.
Use the following steps to calibrate using the initial calibration procedure.
1. Turn the MSI-3460 off.
2. Remove the hex seal screw covering the calibration button (
).
3. Press the
Cal
switch and
simultaneously.
displays.
4. Press
to reset the calibration constants.
displays.
5. Press
to reset.
displays.
6. Press
.
displays.
7. Press
.
will begin flashing.
8. Press
to select lb or kg; when the LED is lit by the correct unit, press
.
displays.
9. Press
. A capacity of 10000 is the initial value and should not be set any higher than load cell rated capacity.
10. Press
to scroll through the numbers and press
to set the number.
11. With the correct number displayed, press
.
displays.
12. Press
. The current scale divisions displays.
13. Press
to scroll through the available scale divisions.
14. When the desired value displays, press
.
displays, follow the standard calibration procedure in
) to complete calibration.
POWER
TARE
TARE
TARE
TARE
USER
F
TARE
TARE
USER
F
TARE
TARE
TARE
USER
F
TARE
Содержание Challenger 3 MSI-3460
Страница 1: ...MSI 3460 Challenger 3 Technical Manual PN 152161 Rev F June 24 2021...
Страница 44: ...MSI 3460 Challenger 3 40 Visit our website www RiceLake com...
Страница 45: ......