Calibration
© Rice Lake Weighing Systems
●
All Rights Reserved
13
17. Press
to select the
value (amount of test weight needed on scale for calibration of
).
18. Press
to accept value.
displays.
19. Place the
weight on the weigh platter and wait for the value to stabilize.
20. Press
to accept and calibrate at
.
21.
briefly displays, followed by the
weight value. Remove
test weights.
22. Press
to select the
value (amount of test weight needed on scale for calibration of
).
23. Press
to accept value.
displays.
24. Place the
calibration weight on the scale and wait for the value to stabilize (the raw A/D counts displays).
25. Press
to accept and calibrate at
. The calibration data is saved and the scale returns to weigh mode.
The weight value displayed must match the value of the calibration weight used. If not, perform the calibration a second time
and follow each step carefully. If
displays, there is a calibration error. Ensure the correct calibration weight value was
selected in comparison to the actual calibration weight used. See
troubleshooting.