3DQHOSULQWHU
DEKO 260 is equipped with a panel printer. It prints cycle report with all basic data and events.
The printer is a thermal printer and the paper report is archiveable.
Cycle report is printed automatically during the process.
There is one button with a led on top right to feed paper
and to indicate the status of printer.
The thermal paper with preservability of 10 years
can be ordered from Franke or it’s representatives.
Cycle print
Install the paper according to the picture above.
Cycle Report:
Owner: PMK Hospital
Machine: 4
Cycle Name: INSTRUMENT
Cycle number:
1452
Cycle start time: 06-08-17
15:45:03
15:45:10 CW rinse start
15:46:34 CW rinse end
15:46:37 Wash start
15:47:26 Chem flow OK
15:47:30 Chem : set 50 ml
T:47
0
C
15:47:35 Temperature attained
T:50
0
C p: 1.05 bar
15:48:30 Wash end
15:48:33 HW rinse start
15:49:30 Temperature attained
T:50
0
C p: 1.04 bar
15:49:30 HW rinse end
15:49:33 Disinfection 90
0
C start
15:55:21 Temperature attained
T: 90
0
C p: 1.01 bar
15:56:26 Disinfection end
T:91
0
C p: 1.00 bar
15:56:30 Disinfection time 60s
15:56:33 Drying start
16:36:34 Drying end
Cycle end time:06-08-17 16:36:45
Cycle passed
-----------------------
signature