Testing and Care
Wander Management Transmitters User Guide
Page 17 of 28
0510-1122-J
TESTING AND CARE
Transmitter
Testing
Do not use the transmitter tester on a
transmitter with an expiration date that has
expired. Results may not be accurate.
All steps are mandatory.
CAUTION
: You must test all transmitters
prior to use to verify proper operation. This
includes every time that the band is
replaced. Failure to test the transmitters
before use can result in system failure
and/or an elopement. In addition, failure to
test transmitters voids the RF Technologies
Product Warranty.
Visual
Inspection
1. Verify that the warranty expiration date
that is stamped on the transmitter is not
expired.
2. Visually inspect the transmitter for
damage or loose parts.
Operation
Test the operation of the transmitter using
the Transmitter Tester. The transmitter
tester will detect whether or not a transmitter
is emitting a signal, but cannot indicate the
strength of the signal.
1. Place the transmitter tester directly on
the transmitter.
2. Press and hold the button on the left
side of the transmitter tester.
3. The device beeps once when you
initially press the button.