Safety Instructions 

IMPORTANT SAFETY INSTRUCTION 

Unplug ST710 from the wall outlet and refer servicing to qualified service personnel under the following conditions. 

In case any abnormality such as smoke, smell or unusual noise are found.

If water or any other liquid is spilled into or onto the ST710.

If power socket or cord is extremely hot.

If ST71 O has been dronned, or the enclosure damaged.

Intended Use 

The SENSATEC ST710 fall monitor, monitors a bed, 
chair, or room for occupancy.  When mounted on a 
bed, chair, or wheelchair, the monitor alarm sounds 
to alert personnel to the risk of a fall with a local 
alarm when a patient/resident is attempting to exit the 
bed, chair, wheelchair, or room without assistance. 
When you hear the alarm, check the patient/resident 
immediately. 

Contraindications 

SENSATEC monitors are not a substitute for 
proper nursing care. The effectiveness of the 
SENSATEC system relies entirely on an 
immediate response by  the caregiver to the 

monitor's alarm. 

Potential Electromagnetic or 

Other Interference 

Take precautions to ensure that the system is 
protected from electromagnetic interference. 
Precautions include: 

• Operate the system at least 15 feet away

from motors, typewriters, elevators, and
other sources of strong electromagnetic
radiation.

• Operation in an enclosed area (wood,

plaster or concrete walls, floors and
ceilings) help prevent electromagnetic
interference.

• Special shielding may be required if the

console is to be operated in the vicinity
of Radio broadcast equipment.

• This medical equipment is approved, in

terms of the prevention of radio wave

interference, to be used in hospitals,

clinics and other institutions which are
environmentally qualified. The use of this
equipment  in an inappropriate environment
may cause electronic interference. This
equipment can be used in residential areas
only.

ST710 Instruction Manual  (0510-0519-

B

Before Using the ST710 

Be sure to read the Instruction Manual thoroughly
before using to get a full understanding of the
ST710.

Be sure to read the following Safety Instructions

carefully. Product specifications described in this

material are subject to change without notice for

reasons of modification and/or improvement.

Keep this manual for future reference.

ST710 is designed to support monitoring patients.

Using a ST710 will not prevent injury, and is not a

guarantee of safety.

Using ST710 in conjunction with facilities' standard

safe practices will help reduce the risk of injury but
is not guaranteed.

Environmental Requirements 

For proper functioning of the ST710, care 
must be taken when it is transported or stored. 

Operational 

Storage 

Transport 

10  to40  C 

-10to60_C 

-10_to60_C 

Temperature 

....... :: ........ :: ...... .......................... .......................... . 

50 to 104  F 

14  to 140  F 

14_ to 140_F 

30 to 75% 

30 to 90% 

Humidity 

non-condensing  non---{Oondensing 

Pressure 

700 to I 060hPa 

700 to 1060hPa 

SAFETY INFORMATION 

30 to 90% 

non-condensing 

700 to 1060hPa 

In these User Instructions, the word WARNfNG 
indicates a potentially hazardous situation which, 
if not avoided, could result in death or serious injury. 
The word CAUTION means a  potentially hazardous 
situation which, if not avoided, may result in minor 
or moderate injury. CAUTION also may be used to 
alert against unsafe practices. 

Terms 

The following terms are used throughout this manual to 
bring attention to the presence of hazards of various risk 
levels, or to important information concerning use of the 
product. 

Page  3  of 11 

Содержание Sensatec ST710

Страница 1: ...ST710 Instruction Manual PN 0510 0519 B Release Date 11 17 17 3125 North 126 th Street Brookfield WI 53005 USA phone 800 669 9946 web www rft com ...

Страница 2: ...3 WARNINGS 4 CAUTIONS 4 ST710 CONTROL UNIT 6 FRONTSIDE 6 INSIDE BATTERY Box 6 USING THEST710 SYSTEM 7 BED MODE 7 CHAIR MODE 7 FLOOR MODE 8 SELECTING A TONE FOR THE ALARM 9 To TURN OFF THEST710 9 DISPOSE OF PROPERLY 9 MOUNTING ATTACHMENTS 9 ALARM INDICATORS 9 TESTING THE SYSTEM AND TROUBLESHOOTING 10 TESTING THESYSTEM 10 TROUBLESHOOTING 10 SPECIFICATIONS 11 Page 2 of 11 ST710 Instruction Manual 051...

Страница 3: ...terference to be used in hospitals clinics and other institutions which are environmentally qualified The use ofthis equipment in an inappropriate environment may cause electronic interference This equipment can be used in residential areas only ST710 Instruction Manual 0510 0519 B Before Using the ST710 Be sure to read the Instruction Manual thoroughly before using to get a full understanding oft...

Страница 4: ...teries Explosion and liquid leak ofbattery may lead to fire and injury Inspect the plug and the outlet regularly Remove the dust and dirt between the plug and the outlet regularly Failure to do so may result in short circuit and fire CDcautions SENSATEC fall monitors are designed to be used with the SENSATEC pressure sensor pads and other accessories to construct a fall monitoring system It is int...

Страница 5: ...g the device for an extended period of time Ifthere is corrosion on the battery terminals clean and wipe offthe battery case and replace with new batteries Ifthe liquid leak comes in contact with body rinse well with water Sensor pad cautions To keep the sensor pads in good condition Do not immerse the sensor pad into liquid ofany ki nd This may cause damage to the product Stop using the sensor pa...

Страница 6: ...Jlljack Alarm Speaker Speaker for alarms Inside Battery Box H M l DllayTime Os 11 2s 4s Sal1ct I I I I L iJ 1 5V AA Battery 1 5V AA Battery 1 5V AA Battery I Page 6 of 11 ST710 Control Unit Inside Battery Box continued Volume Switch Alarm volume switch H M L H High volume M Medium volume L Low volume Delay Time Switch Alarm time delay setting to avoid unnecessary alarm caused by patient s tossing ...

Страница 7: ...h switch to set how quickl y an alarm should go off when 4 receiving the signal from a sensor Set the alarm time delay by sliding the time switch to set how quickly an alarm should go offwhen receiving the signal from a sensor Check ifthe ST710 and sensor is working properly See page 10 Testing the System for detail Check ifthe ST710 and the sensor pad are working properly See page 10 Testing the ...

Страница 8: ... By the bedside Alert when patient tries to leave the bed Plug the floor mat sensor cable into the ST710 sensor jack Doing so automatically places the ST7l0 in Floor Mode Removing the floor mat sensor cable from the ST7l0 turns offFloor Mode operation and returns the ST7lOto Bed Mode Chair Mode Make sure the cable will not interfere with the bed wheels or movable bed components and that no one wil...

Страница 9: ...19 B Mounting Attachments The ST710 can be fitted with the following attachments Strap The strap is attached to the ST710 through a strap hole and can be attached to a bed chair or handrail Mounting Clip The ST710 unit can be attached to the wall or chair through a stainless steel wheelchair clip or wall mounting bracket Strap hole Strap Clip mounting unit I D Chair clip Wall mount clip Alarm Indi...

Страница 10: ...ord is securely plugged in 2 Place weight on the sensor pad Make sure the Red LED starts blinking and the alarm sounds 3 Press the Hold button to stop the alarm 4 Press the Hold button again to reset the alarm Page 10 of 11 Troubleshooting Sensor Pad Cannot Detect Movement When the sensor pad cannot detect the patient s movement because she he is too lightweight Try placing the sensor in another p...

Страница 11: ... Mat Sensor resistance less than 3k ohm Size WxDxH 70mm x 27mm x 113mm Water protection classification IPXO Weight 2 Approx 172g 1 24Hrs Alarm 20sec volume max 5 times day 2 Without protective silicon boot with batteries ST710 Instruction Manual 0510 0519 B Page 11 of 11 Accessories The SENSATEC ST710 monitor can be used with the following SENSATEC sensor pad and mats Sensor Pads RFT l000 1801 30 ...

Отзывы: