SERENA P GB-R04 del 03-07-17
00/Per Bertani/Manuali d’uso Serena/ Serena con stampante
37/99
VACUUM TEST DIAGRAM
==========================================
SERENA 23
ONLY B CYCLE
Ver. 0.27
S/N 000000
Nr.CYCLE 0000
CYCLE: VACUUM TEST
OPERATOR CODE 01
DATE 00-00-00
TIME 00:00:00
PHASE TIME TøC P mbar
VT-0 00:00:16 19.7 992
VT-1 00:03:30 19.6 47
VT-2 00:08:30 20.1 45
VT-3 00:18:30 21.1 45
CYCLE OK
PHASE
DESCRIPTION
VT-0
PRESSURE ENVIRONMENT
VT-1
TIME IN WHICH THE MAXIMUM
DEPRESSION IS REACHED .
VT-2
START TIME COMPENSATION
VT-3
START LEAK PERIOD