Installation and Operation
Blood Bank Refrigerators

7

Operation 

7.1

Start Up

To start up the refrigerator, complete the following steps:

1. Plug in the power cord.
2. Insert the key in the switch and turn the power on.

The temperature display has a range of 0 to 11.4

°

C. When the 

temperature in the cabinet drops below 11.4

°

C, the display 

shows the actual cabinet temperature.

3. Rotate the power switch to the ALARM ON position when 

the temperature drops below the warm alarm setpoint.

4. Allow the unit to reach operating temperature before loading 

it with any product. To stabilize the temperature profile,
a 24-hour waiting period is recommended. 

5. After the unit has pulled down to the desired operating 

temperature, turn the three position key switch one turn 
further clockwise to the Alarm On position.

6. If you have a remote alarm, hook it up at this point (refer to 

Section 9.3 on page 7)

7. Whether you have a built-in alarm or a customer-installed 

remote alarm, you should test it following the instructions in 
Section 9.4 on page 7. Alarm setpoints are factory pre-set for 
5.5 and 1.5

°

C.

7.2

Product Loading Guidelines

When loading your blood bank refrigerator, take care to observe 
the following guidelines:

Never load the blood bank beyond capacity. Table 1 on 
page 1 sho
ws the number of 450ml bags that can safely be 
stored in your unit.

Distribute the load as evenly as possible. Temperature 
uniformity depends on air circulation, which could be 
impeded if drawers are overfilled, particularly at the top of 
the cabinet. 

For critical applications such as blood storage, be sure that 
the alarm systems are working and active before you load 
any product.

7.3

Automatic Defrost

The defrosting process on all models is primarily accomplished 
by air circulated during off-cycle periods. All models are 
equipped with a built-in, on-demand defrost feature which is 
preset at the factory. The units defrost only under the most severe 
conditions.

Under normal conditions, the temperature warm-up during 
defrost is virtually unnoticeable. However, an occasional 2

°

warm-up is possible if usage is heavy and ambient conditions are 
extreme.

8

Temperature Control

8.1

Setpoint Control

Your individual unit has been adjusted and factory tested to 
maintain a +4

°

C cabinet temperature. For blood banks, cabinet 

temperature setpoint is a computed value based on the service 
parameters Cut On and Differential (see Table 4 on page 5). 

For most blood bank applications, however, you will not need to 
change the temperature setpoint.

8.2

Sensor Bottles

Cabinet temperature is measured by probes inserted in sensor 
bottles. The bottles contain equal portions of glycerine and water 
to simulate blood samples. Keep the sensor bottles completely 
full up to the top of the bottle.

Space Saver models have one factory-installed sensor bottle. In 
Upright models (23 ft

3

 and larger) there are two sensor bottles, 

positioned as shown in Figure 4.

Figure 4.  Sensor Bottles (Upright Models)

The default sensor reading for Upright units is the bottle at the 
top of the cabinet. You can check the temperature in the lower 
sensor bottle by pressing and holding 

 and   simultaneously 

(refer to Figure 3 on page 4). When you release the buttons, the 
temperature display returns to the upper sensor reading. 

Содержание REB-1204

Страница 1: ...ining this product please be sure to read this manual and product warning labels carefully Failure to follow these instructions may cause this product to malfunction which could result in injury or damage Below are important safety precautions that apply to this product Use this product only in the way described in the product literature and in this manual Before using it verify that this product ...

Страница 2: ...ed power lines Always connect the equipment to a dedicated separate circuit Electrical codes require fuse or circuit breaker protection for branch circuit conductors Use time delay fuses for 12 AWG circuits For all blood bank models the wiring diagram is attached to the back of the cabinet 4 3 Leveling The unit must be level Rotate the leveling screws located under the front corners of the unit un...

Страница 3: ...g tabs from the slots on the slides Note The drawers fit snugly between the slides Push the back of the drawer from underneath to remove the drawer 3 Raise the back of the drawer almost to a vertical position and disengage the front mounting clips from the slides To remove the drawers on Space Saver models depress the hooks located on both sides of the drawers about 1 3 of the way back and slide t...

Страница 4: ...nts are 8ºC and 0ºC Then the number of bars illuminated indicates cabinet temperature as follows bulb only 0ºC cold alarm 6 bars 4 8 1 bar 0 8 7 bars 5 6 2 bars 1 6 8 bars 6 4 3 bars 2 4 9 bars 7 2 4 bars 3 2 10 bars 8 warm alarm 5 bars 4 setpoint When cabinet temperature exceeds the warm alarm setpoint the top bar of the thermometer flashes When temperature is lower than the cold alarm setpoint t...

Страница 5: ...larm test Test by simulating warm alarm conditions Press and simultaneously hold for five seconds Display and thermometer show simulated cabinet temperature alarms flash and sound as appropriate Display blinks twice when test is completed Lower sensor Display lower cabinet temperature Press and simultaneously and hold Display shows lower cabinet temperature while buttons are held returns to upper ...

Страница 6: ...r critical applications such as blood storage be sure that the alarm systems are working and active before you load any product 7 3 Automatic Defrost The defrosting process on all models is primarily accomplished by air circulated during off cycle periods All models are equipped with a built in on demand defrost feature which is preset at the factory The units defrost only under the most severe co...

Страница 7: ...e purple and red white wires must be tied together in this application The wiring diagram is attached to the inside of the alarm back cover 9 4 Alarm Test Note It is important to test your alarm system after any maintenance operation or temperature control adjustment 9 4 1 Theory of Operation This test procedure applies to both factory installed built in alarms and optional field installed remote ...

Страница 8: ... needs to be changed 10 3 Changing Chart Paper To change the chart paper complete the following steps 1 Locate the pressure sensitive buttons at the front upper left of the recorder panel 2 Press and hold the change chart button for one second The pen will move off the scale 3 Unscrew the center nut remove the old chart paper and install new chart paper Carefully align the day and time with the re...

Страница 9: ...Condensers should be cleaned at least every six months In heavy traffic areas condensers load with dirt more quickly Failure to keep the condenser clean can result in equipment warm up or erratic temperatures CAUTION Never clean near condensers with your fingers Some surfaces are sharp Periodically check the condenser to make sure that it is clean In all models the condenser is located in the top ...

Страница 10: ...mperature continues to fluctuate call an authorized Revco service representative Unit warms up Door is open Make sure the door is completely closed Door seal Check the door seal following instructions in Section 4 5 on page 3 Warm product recently loaded in unit Allow ample time to recover from loading warm product Power supply Check for proper voltage to the unit If there is no voltage to the uni...

Страница 11: ...Can dial up to four telephone numbers to advise of alarm condition across any telephone system which accepts pulse dialing One system can monitor up to four individual cabinets In addition the system can monitor up to three groups of cabinets Contact Revco for detailed specifications Customer installed 6224 Description Catalog No Temperature recorder 6 in circular chart seven day drive free standi...

Страница 12: ...months from the date of shipment by Revco If any part is found to contain a manufacturing defect in material or workmanship Revco assumes no responsibility for any labor expenses for service removal or reinstallation required to replace part or for incidental repairs and such costs are the responsibility of the Owner unless a service labor agreement exists between the Owner and his Dealer The warr...

Страница 13: ...plate attached to the equipment If available please provide the date purchased the source of purchase Revco or specific agent rep organization and purchase order number If You Need Assistance REVCO SALES DIVISION Phone 828 658 2711 800 252 7100 FAX 828 645 3368 LABORATORY PARTS and SERVICE Phone 800 438 4851 FAX 828 658 2576 TECHNICAL SUPPORT Phone 800 438 4851 275 Aiken Road Asheville NC 28804 U ...