DIRECTIONS FOR USE
AIR WHIRLPOOL (OPTION)
FUNCTIONALITY
Fresh clean air is ingested into the system. The air is then compressed passed through a heat-exchanger to heat the air to a
specific temperature. The air is then expelled from the compressor at high pressure and passes into a multiplex-manifold
connected via airlines to 12 de-fuser nozzles positioned specifically along the bottom of the bath. These nozzles aim the
massage jets at a patient’s major muscle groups along their anatomy. The air bubbles explode on impact with the patients
muscles delivering a precise and well direct massage therapy. This stimulates the muscles and builds muscle tone and
strength. The massage therapy also relaxes the patient which is also essential to a patients experience and wellbeing
! NB.
The system cannot be operated unless sufficient water is in the bath tub.
To start hydro-massage system
: press the On/Off button once
To stop the hydro-massage system
: press the On/Off button once again.
+ Increase intensity button:
Press the + button to increase the massage intensity.
- Degrease intensity button:
Press the – button to decrease the massage intensity.
Alternating Massage function:
Press both +&- buttons simultaneously for three seconds to activate the alternating
massage function.
Massage cycle:
The massage cycle is 0-30 minutes. After 30 minutes the system automatically switches off. If the
baths water is emptied during the massage cycle. The system will switch off when the water level reaches the water
sensor when emptying.
Auto-Clean function:
Once the hydro-massage function has been switched off. The system hibernates for a period
of approximately 30 minutes to allow bath to empty and the room temperature to drop reducing the collation of
humidity. This is indicated by the Green On/Off LED light flashing intermittently. After the 30 minute period the
system will automatically turn on and run for a period of 2 minutes. This enables hot air to be blown though the
system of airlines drying any humidity or water ingress making them clean and safe for future use.
Manual Cleaning Cycle:
The system can be cleaned manually whilst water is off and no water is in the bath. To
perform the manual procedure simply press the On/Off button every 2 seconds for two seconds.
Green LED light indication:
Intermittent green flashing light (1s/1s) indicates the presence of water.
Permanent green light- indication system is in “On” mode.
Intermittent green light every (0.5s/0.5s) indicating system is in “Auto-Clean” mode.
Intermittent green flashing light every (1s/7s) - System running without water.
13/21
System On/Off
- Lower Intensity
+ Higher Intensity
Led indicator