8.2 Replacing the machine fuses
Pos : 7. 10 /0030 War nhi nweis e/ W0014 WARNUNG Lebensgef ahr durc h Stroms toß Wec hs el Sic herung Frei li egende Stromkont akt e @ 4\ mod_1332933368704_9.doc x @ 27530 @ @ 1
WARNING
Mortal danger from electric
Exposed power contacts
–
When replacing fuses on the cutout or fuse adapter you may come into
contact with live contacts. An electric shock can lead to burns and to
cardiac arrhythmias or to respiratory arrest and cardiac arrest.
•
Remove the mains
Pos : 7. 11 /0005 RET SCH/ 0005 R ETSCH Bedienungsanlei tungen Kapit elsamml ung en/0005 milling/ XRD-Mill McCrone/Modulordner/0035 Rei nigung und Wart ung XRD-Mill McCrone/ 3510 Modul Sic herung wec hs eln @ 9\ mod_1427186918912_9. doc x @ 76740 @ @ 1
Fig. 16: Replacing fuses
NOTICE
Always replace both fuses.
–
Type of fuse: 2 x T2A 250V
•
Press the two clips (GL) of the fuse holder (GS)
together and pull the fuse holder
•
Remove the fuses from the fuse holder and insert the new fuses.
•
Place the fuse holder with the inserted fuses into the opening
GL
Cleaning and service
Replacing the machine fuses
WARNING
Mortal danger from electric shock
Exposed power contacts
When replacing fuses on the cutout or fuse adapter you may come into
contact with live contacts. An electric shock can lead to burns and to
cardiac arrhythmias or to respiratory arrest and cardiac arrest.
Remove the mains cable before replacing fuses.
Replacing fuses
Always replace both fuses.
Type of fuse: 2 x T2A 250V
Press the two clips (GL) of the fuse holder (GS)
together and pull the fuse holder out.
Remove the fuses from the fuse holder and insert the new fuses.
Place the fuse holder with the inserted fuses into the opening
Cleaning and service
31
W00014
When replacing fuses on the cutout or fuse adapter you may come into
contact with live contacts. An electric shock can lead to burns and to
cardiac arrhythmias or to respiratory arrest and cardiac arrest.
cable before replacing fuses.
Remove the fuses from the fuse holder and insert the new fuses.
Place the fuse holder with the inserted fuses into the opening
GS
GL
Содержание XRD-Mill McCrone
Страница 2: ...2 Copyright Copyright by Retsch GmbH Haan Retsch Allee 1 5 D 42781 Haan Federal Republic of Germany ...
Страница 3: ...3 ...
Страница 34: ...34 ...
Страница 38: ......
Страница 39: ......
Страница 40: ...Copyright Copyright by Retsch GmbH Haan Retsch Allee 1 5 D 42781 Haan Federal Republic of Germany ...