2/25/2013
Retsch GmbH
21
0002
General
Cleaning
TG200
The TG200 should be cleaned only with brushes, and possibly
with an industrial vacuum cleaner or with compressed air.
The system of protection for the TG200 with drying vessel
and clamping lid is IP43.
Without vessel and lid this is reduced to IP22.
Do not clean the TG200 with running water.
Danger to life through current surge.
Filter bags
The filter bags can be washed out under running water.
Filter elements and filter on the machine rear side
These must be replaced when contaminated.
Drying vessels 6 and 0.3 litres
These can be cleaned under running water or in an ultrasonic
bath.
Maintenance
F1
The 3 filters
F1
on the rear side of the machine must be
replaced after approx.
100 hrs.
The maintenance interval is
shortened depending on the dustiness of the surroundings. This
must be assessed empirically by the operator.
If the filter elements and bags are contaminated too much the
efficiency of the TG200 will be reduced to an unknown extent.