2/25/2013
Retsch GmbH
20
0002
Attainable temperatures at the measuring point
Since the attainable temperature depends on the set fan
power, the charged material to be dried and the type of
clamping lid, the table below serves as a guide. The
temperature in the TG200 is measured at the measuring point
TM.
Operating conditions 1:
TG200 without material to be dried, with basic equipment –
clamping lid with filter bag and 6 ltr glass vessel as separate
accesory, Art no. 72.783.0001
TM
Max. temperature set at 150°C or 302°F
Ambient temperature of 23°C / 73.4°F
Time 8 min
Air flow
Attainable
temp °C
(±10%)
Attainable
temp °F
(±10%)
99
ca. 87°C
ca. 188°F
60
ca. 93°C
ca. 199°F
40
ca. 100°C
ca. 212°F
20
ca. 128°C
ca. 262°F
10
ca. 150°C
ca. 302°F
Operating conditions 2:
TG200 without material to be dried and with attachment with 3
glass vessels, each of 0.3 ltr, as separate accessory - Art. No.
72.002.0005
Max. temperature set at 150°C or 302°F
Ambient temperature of 23°C / 73.4°F
Time 8 min
Air flow
Attainable
temp °C
(±10%)
Attainable
temp °F
(±10%)
99
ca. 108°C
ca. 226°F
60
ca. 115°C
ca. 239°F
40
ca. 124°C
ca. 255°F
20
ca. 150°C
ca. 302°F
Operating conditions 3:
TG200 without material to be dried and with as separate
accessory - Art. No. 72.643.0001
Max. temperature set at 150°C or 302°F
Ambient temperature of 23°C / 73.4°F
Time 8 min
Air
flow
Attainable
temp
°C
(±10%)
Attainable
temp °F
(±10%)
99
ca. 88°C
ca. 190°F
40
ca. 98°C
ca. 208°F
30
ca. 106°C
ca. 222°F
20
ca. 120°C
ca. 248°F
10
ca. 150°C
ca. 302°F
Содержание TG200
Страница 24: ......
Страница 26: ......
Страница 27: ......
Страница 28: ...Copyright Copyright by Retsch GmbH Haan Retsch Allee 1 5 D 42781 Haan Federal Republic of Germany ...