2/25/2013
Retsch GmbH
15
0002
Triggering the bimetallic element switch in the
heater
The heater is fitted with a bimetallic element switch.
If a fault occurs this switches off the heater at 160°C/320°F +/-
6%. The fan power is immediately set to maximum and the heater
switched off.
After a cooling cycle has taken place the TG200 must be restarted.
2
5
8
16
Display
2
shows fault code
F2
Temperature is reduced, display
5
Fan power is set to max.
99
, display
8
Red
LED over STOP key
16
lights up
The cooling cycle is started.
The bimetallic element switch is triggered when a temperature >
150°C/302°F has been reached. This can result from too low fan
power and at the same time too high stagnation pressure caused by
the material to be dried. Before restarting the machine the
temperature should be reduced by 5 – 10°C.
Otherwise the heater may be damaged.
Setting the air flow
12
13
14
Reducing:
Key
12
reduces the fan power down to the minimum power step,
10
is indicated in display
13
.
Keep key
12
pressed; after 5 sec. quicker adjustment is possible
Increasing:
Key
14
increases the fan power up to the maximum power step,
99
is indicated in display
13
.
Keep key
14
pressed; after 5 sec. quicker adjustment is possible
It is not possible for the value to drop below
10
or to exceed
99.
Interval – continuous operation
10
11
At the time of switching on the
TG200
interval operation is
switched off. Not during the performance possibly
.
When Interval is
switched on the air flow can no longer be changed.
Switching on interval:
Press key
10
LED over key
10
lights up
Pause for 10 sec with fan switched off; heater is switched off
Operation for 10 sec with reselected fan power; heater is
switched on again
Switching off interval:
Press key
11
LED over key
11
lights up