2/25/2013
Retsch GmbH
13
0002
Operating the TG200
Switching on / off
E
The main switch
E
is located on the right hand side of the rear panel
of the
TG200
Switch on the main switch
ON
lights up in the display
drying mode
000
lights up in the display
temp °C/°F
10
lights up in the display
air flow
LED
interval off
lights up
Two dashes (
- -
) light up in the display
time
The values appear at the first start. If the values are
changed they are retained at the end of the programme
and appear when switching on again and pressing the
START key.
The TG200 is now ready for operation, without interval, for
continuous operation without heating and with minimum fan power.
Starting – interruption - stopping
16a
15
16
Starting:
Press START / key
15
Green LED over key
15
lights up
TG200 starts with specified values.
Red LED over key
16a
flashes as long as a temperature has not
been specified.
Specify temperature, red LED is extinguished, green LED lights
up.
During the drying period the temperature is kept constant within a
tolerance. This means that the heater is switched off when the
specified temperature has been reached and is switched on again
when the temperature drops below the specified value. A
temperature difference of +/- 2°C or +/-50°F is possible here.
Quick start with full power:
Press START / key
15
twice
Green LED over key
15
lights up
TG200 starts with full heating power and fan power for an
indefinite running period
The measured temperature is displayed
Stopping:
Press STOP / key
16
– heater is switched off
Red LED over key
16
flashes
Press STOP / key
16
a second time – everything is switched off
Red LED over key
16
lights up
The complete display except
on
is extinguished
Activate LED displays = press key
15
once
New values can be entered
Press START / key
15
a second time
Function
is carried out as with
Starting
Содержание TG200
Страница 24: ......
Страница 26: ......
Страница 27: ......
Страница 28: ...Copyright Copyright by Retsch GmbH Haan Retsch Allee 1 5 D 42781 Haan Federal Republic of Germany ...