Restek Q-sep 3000 Скачать руководство пользователя страница 7

7

visit 

www.restek.com/QuEChERS

Tube Holder Configuration for Q-sep™ 3000 Centrifuge

The Q-sep™ 3000 centrifuge has been designed to work with 50, 15, and 2mL centrifuge tubes using the appropriate acces-
sories.  The purple tube holders are designed to hold 50mL round or conical bottom centrifuge tubes.  Up to 6, 50mL tubes
can be processed at the same time.* The red tube holders and 4-position inserts are designed to hold 15mL round or con-
ical bottom centrifuge tubes.  Up to 18, 15mL tubes can be processed at the same time.  When processing 15mL tubes, one
of the positions in the tube holder must be left empty to prevent the tubes contacting each other when the holder moves
into the horizontal position. Note that either the left or right position may be left empty, but must be consistent across all
holders to maintain a balanced load. Purple inserts can be used to adapt the red tube holder for use with 2mL round or
conical bottom centrifuge tubes. Up to 24, 2mL tubes can be processed at the same time using these inserts.  

Different sizes of tubes can be processed simultaneously, as long as the balance is properly maintained
(see Balanced Loads below).

Balanced Laods

Your centrifuge must contain a balanced load in order to work properly. To insure that the load is balanced, keep these rules
in mind when inserting test tube samples:
1. Opposing tube holders must be identical. 

2. Opposing tube holders must be empty or loaded with an equal number of equally weighted samples. 

3. If an odd number of samples is to be spun, use a water-filled tube balance with the unpaired one.

NOTE: The Q-sep™ 3000 centrifuge is designed to detect serious out-of-balance situations and will safely terminate an
unbalanced run by cutting the power to the motor. If this occurs, wait for the rotor to stop spinning before unlocking.
Balance the load before restarting the centrifuge motor. 

4. When loading tubes into opposing holders, the holders will be balanced when the tubes are loaded diagonally from one

another. See the illustration below. 

Note: When loading 15mL tubes, the maximum capacity is three (3) tubes per carrier. Keep the lower right side position empty to
prevent tubes from contacting each other when the carrier moves to the horizontal position.

Leave one inner
position empty.

*When using 50mL conical bottom centrifuge tubes, place conical tube inserts into the bottom of the 50mL tube
holder to prevent damage to the centrifuge tubes.

Содержание Q-sep 3000

Страница 1: ...s Service Manual cat 26230 26231 The Q sep 3000 centrifuge complies with all requirements of UL standard 3101 20 Can CSA C22 2 No 1010 1 and Can CSA C22 2 No 1010 2 20 NOTE The programmable run time h...

Страница 2: ...maintenance records PRECAUTIONS 1 Do not attempt to unlatch the centrifuge lid unless the Unlocked indicator light is illuminated see pg 4 2 Avoid solvent contact with centrifuge lid 3 Insure that the...

Страница 3: ...e centrifuge only after the rotor has completely stopped Brushless DC motor requires no routine maintenance cool running Clear lid for safe observation of samples and optical calibration of speed Elec...

Страница 4: ...must be closed Open Emergency Stop Button Allows access to the rotor chamber by disengaging the locking mechanism Entry is only permitted when the rotor is stopped Pressing this button during operati...

Страница 5: ...e cut to the motor but you should not be able to fully turn the latch If it is possible to turn the latch and open the lid while the unit is running contact Restek Corporation or your local distributo...

Страница 6: ...l operation 1 Plug the centrifuge into an approved 115 volt cat 26230 or 230 volt cat 26231 AC 50 60 Hz outlet and press the POWER button 2 Push the OPEN EMERGENCY STOP button verify the rotor is not...

Страница 7: ...perly maintained see Balanced Loads below Balanced Laods Your centrifuge must contain a balanced load in order to work properly To insure that the load is balanced keep these rules in mind when insert...

Страница 8: ...spill occurs in the centrifuge or tube holder thoroughly clean the rotor and both the inside and outside of the tube holder or replace the tube holder if necessary Replacement parts can be found on p...

Страница 9: ...the bottom of the rotor with the pin in the motor shaft see picture The rotor should slide onto the shaft freely and the rotor should be level Be sure that the slot in the underside of the rotor lines...

Страница 10: ...started spinning it may be possible to rotate the lid knob enough to cause the pawl to lose contact with the lid safety switch If this happens the centrifuge motor may lose power but the lid will stil...

Страница 11: ...authorized distributor for assistance The printed circuit board may be damaged Contact Restek Corporation or your distributor for assistance The Q sep 3000 centrifuge is equipped with severe out of ba...

Страница 12: ...and workmanship for 5 000 cycles or 12 months whichever comes first Should the centrifuge require warranty or out of warranty service please contact Restek Corporation or your local distributor Lit C...

Отзывы: