Q-sep™ 3000 Supplied Equipment*

The following items come standard with each Q-sep™ 3000 centrifuge:
1. One (1) six place horizontal rotor
2. Six (6) 50mL single-place tube holders
3. Six (6) 50mL conical tube inserts
4. Six (6) 15mm four-place tube holders
5. Twenty four (24) 2mL tube adapters
6. One (1) Operator’s Manual

*The rotor and rotor accessories are rated for a rotation frequency of 4,130 rpm,
(a force of 3000 xg with the 15mm tube holders).

Q-sep™ 3000 Available Features 

Swing-out horizontal rotor design incorporating a unique test tube holder that produces 

horizontally separated samples

while requiring no additional parts. 

Fixed-speed of 4,130 rpm (3,000 xg). 

Cool-flow air flow design that prevents overheating of samples.

Heavy gauge steel construction for safety and durability.

Lid safety switch that prevents the centrifuge from operating unless the lid is closed and latched.

Removable rotor for easy cleaning.

Locking lid that allows entry into the centrifuge only after the rotor has completely stopped.

Brushless DC motor; requires no routine maintenance, cool running.

Clear lid for safe observation of samples and optical calibration of speed.

Electronically controlled timed operation, variable from 1 to 30 minutes.

Push-button operation.

Indicator lights: 

“RUN”

green

-lights when power is applied to the motor 

“LATCH”

yellow

-lights when the lid is closed and latched 

“UNLOCKED”

red

-lights when the lock system is deactivated 

Cycle counter to monitor machine use.

Audible indicator at the end of each run.

1.

2.

3.

4.

6.

5.

Q-sep™ 3000 Specifications

Motor Speed and Force Rating:

4,130 rpm, 3,000 xg

Maximum Capacity with 6-Place Horizontal Rotor:

6 x 50ml tubes, 18 x 15ml tubes, 24 x 2ml tubes

Motor:

1/2 H.P. brushless DC 

Nominal Acceleration Time:

45 seconds 

Nominal Braking Time:

120 seconds 

Timer (electronic):

1 to 30 m/-1% 

Requirement: 

2.0 or 1.0 amps 

Current Voltage Requirement:

115 or 230 (+/-10%) volts 

Frequency:

50 / 60 Hz 

Centrifuge Protection Breaker:

4 amp re-settable 

Overall Dimensions (H x W x D):

9 in (22.9cm) x 14.5 in (36.8cm) x 17 in (43.2cm) 

Weight:

39 lbs (17.7kg) 

See page 11 for replacement part catalog numbers and ordering information.

Содержание Q-sep 3000

Страница 1: ...s Service Manual cat 26230 26231 The Q sep 3000 centrifuge complies with all requirements of UL standard 3101 20 Can CSA C22 2 No 1010 1 and Can CSA C22 2 No 1010 2 20 NOTE The programmable run time h...

Страница 2: ...maintenance records PRECAUTIONS 1 Do not attempt to unlatch the centrifuge lid unless the Unlocked indicator light is illuminated see pg 4 2 Avoid solvent contact with centrifuge lid 3 Insure that the...

Страница 3: ...e centrifuge only after the rotor has completely stopped Brushless DC motor requires no routine maintenance cool running Clear lid for safe observation of samples and optical calibration of speed Elec...

Страница 4: ...must be closed Open Emergency Stop Button Allows access to the rotor chamber by disengaging the locking mechanism Entry is only permitted when the rotor is stopped Pressing this button during operati...

Страница 5: ...e cut to the motor but you should not be able to fully turn the latch If it is possible to turn the latch and open the lid while the unit is running contact Restek Corporation or your local distributo...

Страница 6: ...l operation 1 Plug the centrifuge into an approved 115 volt cat 26230 or 230 volt cat 26231 AC 50 60 Hz outlet and press the POWER button 2 Push the OPEN EMERGENCY STOP button verify the rotor is not...

Страница 7: ...perly maintained see Balanced Loads below Balanced Laods Your centrifuge must contain a balanced load in order to work properly To insure that the load is balanced keep these rules in mind when insert...

Страница 8: ...spill occurs in the centrifuge or tube holder thoroughly clean the rotor and both the inside and outside of the tube holder or replace the tube holder if necessary Replacement parts can be found on p...

Страница 9: ...the bottom of the rotor with the pin in the motor shaft see picture The rotor should slide onto the shaft freely and the rotor should be level Be sure that the slot in the underside of the rotor lines...

Страница 10: ...started spinning it may be possible to rotate the lid knob enough to cause the pawl to lose contact with the lid safety switch If this happens the centrifuge motor may lose power but the lid will stil...

Страница 11: ...authorized distributor for assistance The printed circuit board may be damaged Contact Restek Corporation or your distributor for assistance The Q sep 3000 centrifuge is equipped with severe out of ba...

Страница 12: ...and workmanship for 5 000 cycles or 12 months whichever comes first Should the centrifuge require warranty or out of warranty service please contact Restek Corporation or your local distributor Lit C...

Отзывы: