Restek Q-sep 3000 Скачать руководство пользователя страница 5

5

visit 

www.restek.com/QuEChERS

Initial Setup Procedure

If any problems are found during the initial setup procedure, refer to the troubleshooting section on page 11.

1. Plug the power cord into the rear of the centrifuge. Plug the other end into an approved electrical outlet.

For electrical safety, the unit must always be properly grounded.

2. Engage the power button on the front of the centrifuge (lower left-hand side).

The button will illuminate when it is activated.

3. For safety purposes, the locking system is always activated. To deactivate the system (in order to insert or retrieve sam-

0

ples) press the “OPEN / EMERGENCY STOP” button on the control panel. The “UNLOCKED” indicator light should illumi-
nate. If it does not, refer to page 11 on troubleshooting. The lid will be unlocked for 15 seconds after pushing the
“OPEN/EMERGENCY STOP” button. 

4. Turn the latch counter-clockwise and open the lid. 

5. Spin the rotor by hand; check for free and level rotation. If the rotor does not spin freely,

refer to page 11 on troubleshooting. 

6. Check the thumb screw in the middle of the rotor and make sure that it is tight. 

7. Place the test tube holders inside the rotor and verify that they are seated properly. 

8. Close the lid. Rotate the lid knob clockwise to its complete stop position. The ’LATCH’ indicator light should be illumi-

nated. If it is not, make sure that the lid is latched properly. The centrifuge will not run unless the lid is latched and the
“LATCH” light is on. 

9. Turn the centrifuge on by pushing the “START” button. 

10. The “RUN” indicator light will illuminate. 

11. The unit will accelerate to its set speed. 

12. Listen to the sound of the centrifuge. A smooth whirring sound should be heard. If there are any loud or unusual 

sounds, stop the centrifuge by pushing the “OPEN / EMERGENCY STOP” button immediately and refer to page 11 on 
troubleshooting. 

13. While the machine is running, try to turn the latch counter-clockwise. Power may be cut to the motor but you should

not be able to fully turn the latch. If it is possible to turn the latch and open the lid while the unit is running, contact 
Restek Corporation or your local distributor. Close and latch the lid.

14. Push the “OPEN / EMERGENCY STOP” button. The “RUN” indicator light should go out and the motor should slow to

a stop. 

15. The lid should remain locked until the rotor has stopped. If the machine unlocks prematurely, contact Restek 

Corporation or your local distributor. Once the rotor has stopped, a beeper will sound and the interlock system will 
become disengaged for sixty (60) seconds. The “UNLOCKED” indicator light will be illuminated during this time. 

16. To gain entry into the centrifuge after this period has ended, simply press the “OPEN / EMERGENCY STOP” button.

The lid will unlock for fifteen (15) additional seconds. Repeat if necessary.

Setup Location 

1. Unpack the centrifuge and verify that all of the supplied equipment is present. 

2. Choose a setup location which meets the following criteria: 

a) A bench top clearance height of 24” (60cm) is required in order to open the lid. 
b) The clearance envelope is the space around the centrifuge which is required for safety. Choose a setup location 

which will allow for a clearance envelope of at least 30” (76cm) x 30” (76cm), with the centrifuge at the center. No
person or hazardous material should be permitted in the clearance envelope during operation. The operator time
within the envelope should be limited to the time necessary for loading, unloading, and centrifuge operation only. 

c) Proper ventilation is necessary to prevent the overheating of samples as well as premature failure of the cen-

trifuge. Choose an area which will allow unencumbered air flow. 

d) No adjustment is necessary for leveling the centrifuge; however, the surface should be flat and level.
e) Be sure the outlet is always within reach as the power cord is only the means of emergency disconnection!

After the centrifuge has passed this procedure, it is ready for operation. 

If you would like to make adjustments to your machine’s settings, please continue on to “Additional Set-Up Procedures”. 

Содержание Q-sep 3000

Страница 1: ...s Service Manual cat 26230 26231 The Q sep 3000 centrifuge complies with all requirements of UL standard 3101 20 Can CSA C22 2 No 1010 1 and Can CSA C22 2 No 1010 2 20 NOTE The programmable run time h...

Страница 2: ...maintenance records PRECAUTIONS 1 Do not attempt to unlatch the centrifuge lid unless the Unlocked indicator light is illuminated see pg 4 2 Avoid solvent contact with centrifuge lid 3 Insure that the...

Страница 3: ...e centrifuge only after the rotor has completely stopped Brushless DC motor requires no routine maintenance cool running Clear lid for safe observation of samples and optical calibration of speed Elec...

Страница 4: ...must be closed Open Emergency Stop Button Allows access to the rotor chamber by disengaging the locking mechanism Entry is only permitted when the rotor is stopped Pressing this button during operati...

Страница 5: ...e cut to the motor but you should not be able to fully turn the latch If it is possible to turn the latch and open the lid while the unit is running contact Restek Corporation or your local distributo...

Страница 6: ...l operation 1 Plug the centrifuge into an approved 115 volt cat 26230 or 230 volt cat 26231 AC 50 60 Hz outlet and press the POWER button 2 Push the OPEN EMERGENCY STOP button verify the rotor is not...

Страница 7: ...perly maintained see Balanced Loads below Balanced Laods Your centrifuge must contain a balanced load in order to work properly To insure that the load is balanced keep these rules in mind when insert...

Страница 8: ...spill occurs in the centrifuge or tube holder thoroughly clean the rotor and both the inside and outside of the tube holder or replace the tube holder if necessary Replacement parts can be found on p...

Страница 9: ...the bottom of the rotor with the pin in the motor shaft see picture The rotor should slide onto the shaft freely and the rotor should be level Be sure that the slot in the underside of the rotor lines...

Страница 10: ...started spinning it may be possible to rotate the lid knob enough to cause the pawl to lose contact with the lid safety switch If this happens the centrifuge motor may lose power but the lid will stil...

Страница 11: ...authorized distributor for assistance The printed circuit board may be damaged Contact Restek Corporation or your distributor for assistance The Q sep 3000 centrifuge is equipped with severe out of ba...

Страница 12: ...and workmanship for 5 000 cycles or 12 months whichever comes first Should the centrifuge require warranty or out of warranty service please contact Restek Corporation or your local distributor Lit C...

Отзывы: