background image

 
 

 

 

FREQUENTLY ASKED QUESTIONS (continued)

 

 

When I am in the Concentrator Mode and hook the unit up to a U.S.P.  
oxygen(

>

 99.0%) source the LCD does not read 99% (± 2), why? 

When the unit is in its default (concentrator) mode the device's 
software enables the unit to measure the output of an oxygen 
concentrator by compensating for the argon that passes through the 
sieve beds.  In this mode, it will display a false reading (usually ± 
90%) if connected to an oxygen source other than a concentrator.  
Pressing and holding the 

“pure O 

2

/calibrate”

 button removes the 

argon compensation and allows the device to measure the oxygen 
purity with ±2% of compressed gaseous or liquid oxygen.  

 

When I am in the Concentrator Mode and hook the unit up to a 
U.S.P. oxygen (> 99.0%) source the LCD the display reads - 
"Concentrator O

2  

91.8" (approx).  What's wrong? 

Nothing.  When in the Concentrator Mode, the electronic 
components of the indicator automatically compensates for the argon 
found in the gas produced by an oxygen concentrator.  As a result 
the display will read approximately 90% unless this function is over-
ridden by pushing the 

“pure O

2

/calibrate”

 button. 

   

Whenever I check concentrators with the Respond 

O

2

, I 

consistent get an O

2

 concentration reading of 96.0 on every 

concentrator.  Why? 

Concentrators work at their maximum efficiency (96.0%) at lower 
flow rates.  When the flow rate is increased the output of the 
concentrator typically will decrease slightly.  If you question whether 
the device is functioning properly, stabilize it to room air and then 
retest with the concentrator set on the maximum flow rate or beyond.  
Be sure to allow the concentrator to stabilize at the higher flow rate 
before testing.  This may take as long as 10 minutes. 

 

Will the unit lose calibration when I change batteries? 

No.  The unit holds calibration even when the battery is removed. 

 
 
 
 
 
 
 
 
 

                

12

 

Ultrasonic Oxygen Indicator 

 
 

 

 

FREQUENTLY ASKED QUESTIONS

 

 

The following are some of the frequently asked questions about the 
operation of this device.  

 

Do I have to calibrate my Respond 

O

2

™ 

 

unit daily?  

No. The 

Respond 

O

2

 

indicator was calibrated at the factory.  

 

How do I know that my Respond 

O

2

™ 

device is accurately  

measuring concentrations?  

To verify, simply connect the unit to a source of either compressed 
gas or liquid U.S.P. oxygen (> 99.0%) at 2 LPM.  Then turn on the 
device and hold down the 

“pure OB

2B

/calibrate”

 button. As long as 

the display reads between 

99% and 101%

, the unit is properly 

calibrated.  

 

How often should I calibrate my Respond 

O

2

™ 

 

unit?  

As long as you can verify 

(via the Pure Oxygen Verification 

Procedure) 

that the unit is holding calibration against a U.S.P. 

oxygen (> 99.0%) source, the unit does not require calibration.  

 

My accreditation organization as well as our internal policies 
require periodic calibration of this type of equipment. Can I 
calibrate the Respond 

O

2

 

device to comply with these 

requirements? 

 

Yes, if accreditation organizations or internal policies and procedures  
require routine calibration, instructions on calibrating the device are  
found on

 pages 6 - 7 

of this manual. 

  

Does the unit have to be turned off before verifying or 
calibrating? 

 

Yes. The unit must be turned off and evacuated to room air before 
you begin the oxygen verification or calibration procedures.  

 

Why doesn't the unit drop immediately to room air? 

Oxygen is heavier than room air, so it will linger inside the unit.  
When  
there is a flow (either from room air or another concentrator), the 
reading will change immediately. 

 

Can I exhale (blow) into the unit to reach room air? 

No.  You may apply a small suction or flow of room air, but do not 
exhale into the unit. 

 

 11 

Ultrasonic Oxygen Indicator 

Содержание Respond O2

Страница 1: ...e 866 333 4030 Fax 866 333 4035 sales respondo2 com www respondo2 com 803350 Rev B October 2020 Respond O2 Ultrasonic Oxygen Indicator Operation Manual Responsive Respiratory Inc 6244 Lemay Ferry Rd S...

Страница 2: ...turn the unit to Responsive Respiratory Responsive Respiratory will be responsible for normal shipping charges for the return of the unit to purchaser All warranties are voided if the Respond O2 unit...

Страница 3: ...ration from any source other than a conventional oxygen concentrator using molecular sieve beds This unit will not withstand fluid spillage The unit will not withstand mechanical shock or vibration NO...

Страница 4: ...ies Clean battery terminals When unit is turned on display flashes PRO O2 CAL and then turns off Unit requires calibration Perform Full Calibration Procedure outlined in Operation Manual Display reads...

Страница 5: ...of an oxygen concentrator adjusting for the argon that passes through the sieve beds 2 When using the unit with the pure O2 calibrate button depressed the software measures oxygen not adjusting for th...

Страница 6: ...s Will the unit lose calibration when I change batteries No The unit holds calibration even when the battery is removed 12 Ultrasonic Oxygen Indicator FREQUENTLY ASKED QUESTIONS The following are some...

Страница 7: ...t supply tubing to the concentrator oxygen outlet and the inlet port 5 Turn on the device See above 6 The indicator will stabilize in about 10 seconds and provide an accurate reading on the LCD displa...

Страница 8: ...ircuit will compensate for variations in temperature The temperature during storage and shipping should not reach below 29 F 34 C or above 160 F 71 C If the unit has been stored at other than normal r...

Страница 9: ...cator FULL TWO STEP CALIBRATION PROCEDURE The following procedure may be used to calibrate the indicator to both room air and U S P oxygen 99 0 1 Do not connect unit to oxygen source or introduce oxyg...

Страница 10: ...the rear of the unit away from the case Connect the 9 volt battery to the standard 9 volt battery connection and lay the battery sideways in the battery compartment Insert battery door by sliding door...

Отзывы: