VERIFICATION & CALIBRATION INSTRUCTIONS
PURE OXYGEN VERIFICATION PROCEDURE
To verify that the indicator unit is accurately measuring the concentration of
oxygen being produced by a concentrator, perform the following steps:
1. Connect the inlet port on the unit to a USP oxygen source (cylinder or
liquid oxygen) via a supply tube.
2. Turn the O
2
delivery flow to 2 LPM.
3. Press the
“on/off”
button to turn unit on.
4. When unit stabilizes the display will read -
"Concentrator O
2
91.8"
(approx.concentration.).
NOTE: When in the Concentrator Mode, the electronic components of the
indicator automatically compensate for the argon
found in the gas produced by an
oxygen concentrator. As a result the display will read
90%
(± 2%) unless this
function is over-ridden by pushing the "
pure O
2
/calibrate
" button).
5. Press and continue to hold down the
“pure O
2
/calibrate”
button.
6. Unit will display
“Pure O
2
”
and a numeric reading.
7. If the unit is properly calibrated, the display should read between
99% and 101%
. The
Respond
O
2
™
is ready to use.
8. If the unit display reads
"err, "---", "-U-"
or any other reading
other than 99.9 (± 2%), the unit should be recalibrated (See below).
QUICK “ONE STEP” PURE OXYGEN CALIBRATION
The following procedure can be used to calibrate the
Ultrasonic Oxygen
Concentrator Indicator
to pure oxygen:
1. Follow steps 1 through 6 of "Pure Oxygen Verification Procedure"
above.
2. When the unit displays
"Pure O
2
"
and a numeric reading,
continue holding down the
“pure O
2
/calibrate”
button.
3. After a few seconds, the LCD display (while continuing to read
"Pure O
2
"
) will start to alternately display
"CAL"
and number.
4. Continue to hold the
“pure O
2
/calibrate”
button down until
the display stops alternating and simply displays
“Pure O
2
100”
.
5. At that point, the unit is calibrated to pure oxygen and ready for
use.
NOTE: If unit displays an
O
2
concentration reading outside of the ± 2% tolerance range
(eg. < 98% or >102%) unit will not “Quick Cal” and the Full Two Step procedure on
the following page must be followed.
6
Ultrasonic Oxygen Indicator
FULL "TWO STEP" CALIBRATION PROCEDURE
The following procedure may be used to calibrate the indicator to both room air
and U.S.P. oxygen (> 99.0%):
1.
Do not
connect unit to oxygen source or introduce oxygen to the
unit at this time.
2. Turn unit on by pushing
“on/off”
button.
3. Wait for the unit to stabilize at room air - and the LCD displays
“Concentrator O
2
20.9%”.
If necessary, airflow will purge the
oxygen. Do not continue to flow air during the rest of this
calibration process.
NOTE: Do not exhale breath into unit. Moisture will affect calibration; a slight
suction can be used to evacuate the sampling chamber.
4. When unit is stabilized at room air and displays
“Concentrator
O
2
20.9%”
(± 2%) - turn unit OFF by pushing
“on/off”
button.
5. Next, hold the
“pure O
2
/calibrate”
button down while pressing
and releasing the
“on/off”
button.
6. Continue to hold down the
“pure O
2
/calibrate”
button until the
display reads
“---”
.
7. Release the
“pure O
2
/calibrate”
button and the display will read
“CAL AIR”
.
8. Press and release the
“pure O
2
/calibrate”
button again
9. The unit's display will roll for approximately one minute while it
calibrates to room air.
10. When the unit display reads
“CAL O
2
”
- connect the inlet port
to an oxygen source (cylinder or liquid) using the supplied tube.
11. Introduce U.S.P. oxygen (> 99.0%) at 2 LPM to the unit.
(Do not
use a humidifier).
12. Press and release the
“pure O
2
/calibrate”
button again.
13. The unit's display will roll for approximately one minute while it
calibrates to oxygen.
14. When the calibration sequence is complete the unit will display
“CAL END”
.
15. The unit is then calibrated and ready for use.
16. To verify that the unit is properly calibrated, repeat the "Pure Oxygen
Verification Procedure” on
page 6
.
NOTE:
If unit displays
"CAL ERR"
during verification or calibration
procedures, check to ensure that you are using U.S.P. oxygen (> 99.0%) and that the
flow is turned on to at least 2 LPM, then repeat procedure.
NOTE: If unit displays "---" during verification or calibration procedures, release the
"pure
O2/calibrate”
button or the unit will shut off automatically.
If any problems are encountered during the calibration sequences,
contact Responsive Respiratory at 1-866-333-4030.
7
Ultrasonic Oxygen Indicator